readings currentlystored in the demand or auto log memory. The COUNT pa-
rametershows the number of auto log readings remaining during the current
log session.
Press the Enter or Exit buttontoexitthisdisplayandtoreturntotheAUTO
LOG menu.
Only thosechannels that are enabled during the log session are logged. Al-
though you may enableand disable channels during a log session we do not
recommend this practice. If channels are disabled during the log session, the
previouslylogged entries for those channels are not removed from the log.
From the main display,there are two ways to tell if data is being logged: (1) a
loggingindicator in the lower left-hand corner (looks like a strip-chart) and (2)
by setting one of the display fields to the ALOG type (see Section 8.3.3, Edit
Fields)
7.4.2.3 View Data
The VIEW DATA function displays the label,date, reading number, channel,
measurement, units, and time of each auto log entry in sequential order.When
thisfunction is selected, the bottom portion of the display shows the auto log
data.
Each time theview data function is accessed the current auto log memory is
displayed. The data is updated each time the function is accessed; it is not up-
datedcontinuously during a log session.
The entriesare grouped by the label specified and displayed in the order that
theywere logged. For each log session, an entry is written into the log, which
includesthe label for all readings in that session and the date of the session.
1529 Chub-E4 Thermometer Readout
User’s Guide
62
1
24.159C
3
24.142C
2
24.634C
4
24.015C
DATA_01 10-12-00
0002:
0003:
0004:
12:23:24
12:23:24
12:23:24
24.147 C
29.637 C
23.972 C
1
2
3
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure36 Auto Log View Data