8.4.1.5 FORMat:STAMp?
Returnsthe extended measurement data format setting. See Section 9.4.1.4,
FETCh? for details.
FORM:STAM?
0
8.4.1.6 FORMat:STAMp<bool>
Enables extendedmeasurement data format with the FETCh?, MEASure? And
READ?commands. On powerup this is always set to off.
FORM:STAMON
8.4.1.7 MEASure? [<chn>]
Returns the most recent measurement for the specifiedchannel.
MEAS? 1
0.0127
This command is the sameas the FETCh? [<chn>] command
8.4.1.8 READ? [<chn>]
Returns the most recent measurement for the specifiedchannel.
READ? 2
0.0113
This command is the same as the?FETCh [<chn>] command.
8.4.1.9 SENSe<chn>:DATA?
Returnsthe input value for the specified channel.
SENS3:DATA?
100.0291, 0.0000
The SENSe suffix,<chn>, specifies the channel (1 to 4). The parameter is
ohms for lowrange PRTs, Kohms for high range PRTs and thermistors, and
millivoltsfor thermocouples. The second returned parameter is always 0 for
PRTsand themistors and the reference junction temperature for thermocouples.
8.4.2 Measurement Control CommandsThe measurementcontrol commands are used to query and set the moving av-
erage filter,the resistance range, the number of probe wires, and the sequence
timer.The measurement control commands for resistance range and the number
of probe wires are passwordprotected.
1529 Chub-E4 Thermometer Readout
User’s Guide
88