0
If thisquery returns a 0, an auto log session is not running. If this query returns
a 1, anauto log session is running.
8.4.7.10 LOGging:AUTomatic:STATus<bool>
Sets theauto log status.
LOG:AUT:STAT0
The <bool>parameter turns auto logging on (1 or ON) or off (0 or OFF). If si-
multaneous scan mode is enabled, one measurement is stored for each channel
in the auto log. The *RST command set the auto log status to off.
8.4.7.11 LOGging:AUTomatic:TIMe? [MIN|MAX|DEF]
Returns the auto log interval setting.
LOG:AUT:TIM?
1
Specifying the MIN parameter returns a value of 0.1. Specifying the MAX pa-
rameter valuereturns value of 3600. Specifying the DEF parameter returns a
valueof 1. The units are seconds.
8.4.7.12 LOGging:AUTomatic:TIMe<num>|MIN|MAX|DEF
Sets the auto log intervalsetting .
LOG:AUT:TIM1
The <num>p arametercan be one of the following values: 0.1, 0.2, 0.5, 1, 2, 5,
10, 30, 60, 120, 300, 600, 1800 or 3600. The unitsare seconds. If a value other
than one listed is used, the instrument automatically selects the nearest accept-
able valuethat is less than the value entered. Values greater than 3600 or less
than 0.1 are ignored. Specifying theMIN, MAX o r DEF parameter sets thelog
intervalsetting to the minimum, maximum, or default value respectively.
8.4.7.13 LOGging:AUTomatic:VALue? <num>|MIN|MAX|DEF
Returns thespecified auto log entry.
LOG:AUT:VAL?10
,1,22.860,C,12,19,46,,,
The <num> parameter specifiesthe auto log entry to return. Specifying either
the MINor DEF parameter returns the first entry in the auto log. Specifying the
MAX parameter returns the lastentry in the auto log.
Note: Everytime an auto log sessio n isstarted , the firstentry in the log for that
session consistsof the data label and the date. All subsequent entries in the auto
log for that session consist of the channel, measurement value, units and time.
1529 Chub-E4 Thermometer Readout
User’s Guide