CALC:CONV:PRIN2
“PRB”,CHAN2,”ser”,090105,”TYP”,RES,”RANGE”,0
If no parameters are specified,the ALL parameter is assumed and the parame-
tersfor allchannels are printed to the current port. Specifying the <chn> pa-
rameter printsonly the parameters for the specified channel (1 to 4) to the
current port. Optionally,the parameters may be printed to a specific port by
specifying the<port> parameter (see Table 10 on page 125, Port Numbers)
8.4.4.9 CALCulate<chn>:CONVert:SNUMber?
Returnsthe probe serial number for the specified channel.
CALC2:CONV:SNUM?
A_336C
The CALCulatesuffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to14). The returned serial number is in string format.
8.4.4.10 CALCulate<chn>:CONVert:SNUMber <serl>
Sets the probe serialnumber f or the specifiedchannel.
CALC2:CONV:SNUMA_336C
The CALCulatesuffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14).The <serl> parameter is in string format. The <serl> parameter
can consistof u p toeight characters that include any letters, numeric digits, and
the underscore ‘_’.
This command may be password-protected. Toaccess this command:
If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN<pass> command (see Section 9.4.11, Password Com-
mands).
8.4.4.11 CALCulate<chn>:CONVert:TEST? <res>|<volt>
Teststhe temperature conversion for the specified channel.
CALC2:CONV:TEST?100.0145
0.0100
The CALCulatesuffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14).For PRT/Thermistor channels, the <res> parameter specifies a re-
sistance valueto convert to temperature. The <res> parameter is always in
Ohms for PRTsand KOhms for Thermistors. For Thermocouple channels, the
<volt>p arameter specifiesa voltage value to convert to temperature. The
1529 Chub-E4 Thermometer Readout
User’s Guide
96