The labelsuffix, <n>, specifies the number of the label (1 to 25) to retrieve.
The labelname is returned in string format.
8.4.7.23 LOGging:LABel<n>:NAME <label>
Sets the name of the specified label.
LOG:LAB18:NAMEBATH
The labelsuffix, <n>, specifies the number of the label (1 to 25) to set. Label
names can be up to8 characters in length and can include any letters, numeric
digits,and the underscore (‘_’).
8.4.8 System CommandsThe systemcommands are used for querying instrument identification parame-
ters, temperature units, and for setting operating parameters to the defined de-
faultvalues.
8.4.8.1 *IDN?
Returns the instrument identification string that indicates the manufacturer,
model number,serial number, and code firmware version.
*IDN?
HART,1529,A09001,1.11
8.4.8.2 *OPT?
Returnsa list of input modules installed.
*OPT?
PRT,TC
Input module typesare reported b y position, top to bottom,of the two input
modules. Thereturned example is for a 1529 equipped with one 2-channel PRT
inputmodule and one 2-channel Thermocouple input module.
8.4.8.3 *RST
Sets the instrument operating parameters to defined conditions.
*RST
The following parameters are set when this command is issued.
•Sets the scanning sequence timer to 1 second.
•Sets channel 1 ON and all others OFF.
•Sets the scan mode to SCAN.
•Sets the moving average to 1.
•Sets the printing to all ports OFF.
109
8 Digital Communications Interface
Commands