of the displayshows the current display settings. This function is the same as
the FIELDS menu DISPLAYOPTIONS function.
The RESOL parameter determines the number of decimal places (0 through 4
or AUTO)to be displayed for the temperature measurements. If AUTO is se-
lected, the instrument automatically determines the number of decimal places
to displaydepending on the probe conversion type.
The DECIMAL parameter is for selecting thecharacter used as the decimal
separator.The choices are period (.) or comma (,).
The LAMP parameter specifiesthe brightness of the display (OFF,LOW,
MED, HIGH).Enabling this parameter causes the battery to discharge faster.
Use theLR buttons to change the parameter setting. Use the UD buttons
to scroll between thep arameters. Pressthe Enter buttontosavethenewset
-
ting.Press and hold the Exit button to cancel to the main display or press the
EXIT buttonto cancel and to move to the next parameter.
When the AUTO-CALfunction is selected, the following message is displayed.
FASTMODE AUTO CAL.
Press ENTER to continue...
Press the Enter buttonto perform the full measurement function. Press the
Exit buttontocancelandtoexittothemenu.
7.2 Probe MenuThe PROBEmenu provides functions for specifying the channel, probe type,
the typeof temperature conversion and coefficients, copying probe parameters,
33
7 Menu Functions
Probe Menu
1
24.159C
3
24.142C
2
24.634C
4
24.015C
RESOL: 3
DECIMAL:
.
LAMP: OFF
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure12 Display Options