6.8.2 Enabling the Channel

Channels are set firstby selectin g either simultaneous or scan mode in the
CHANNEL menu CHANNEL MODEfunction. Simultaneous mode shows one
to four channelsin the upper display simultaneously. Scan mode shows one
channel ata time (in large digit format) scanning through one to four channels
depending on the channels thatare enabled. After the mode has been selected,
use the CHANNEL menu ENABLECHANNEL function, to select the channel
to be displayed.Any channel set to ‘Off ’is not displayed. See Sections 7.1.3,
ChannelMode, and 7.1.2, Enable Channel, for more information.

6.8.3 Selecting Conversion Type and Probe Characterization

Before the 1529 can accurately measure temperature, it must be configuredto
calculate temperature from the resistanceor voltageof the sensor. There are
manytemperature conversion algorithms available and the one to use depends
on the typeof sensor and its calibration. Many conversion algorithms use coef-
ficientsthat characterize the sensor. Coefficients are determined when the sen-
sor is calibrated.SPRTs and RTDs often use the ITS-90 algorithms and are
providedwith ITS-90 characterization coefficients. Thermistors often use the
Steinhart-Hartalgorithms and coefficients. Thermocouples use standard tables
or equationsdepending on its type. For additional information on conversion
types, see Section7 .2, Probe Menu.
The conversiontype and characterization coefficients for a sensor are specified
using the PROBEmenu EDIT PROBE function (see Section 7.2.1).
6.9 Fast Measurement Mode
FastMeasurement Mode applies to measure periods of 0.1, 0.2, or 0.5 seconds.
This modeallows measurements to be displayed or logged quickly.
Normally,when each measurement is made a self-calibration of the measure-
ment circuitis performed simultaneouslyduring the measurement process to
offseterrors from component drift and spurious EMFs. For measure periods of
0.1, 0.2, and 0.5 seconds, thefast measurement rate is achieved by foregoing
the self-calibration. The drawbackto this FastMeasurement Mode is the accu-
racyof the measurement may be poor and subject to drift. Accuracy can be
temporarilyimproved using the AUTO-CAL function that appears in the
CHANNEL menu when in thisFast Measurement Mode. The AUTO-CAL
function causes the instrument to perform a single self-calibration of the mea-
surement circuitthen resume fastmeasuring (without automatic
self-calibration).
See Section 8.1.1 for detailed information on using the Fast Measurement
Mode.
1529 Chub-E4 Thermometer Readout
User’s Guide
24