SectionI

Model34904

(Cont'd).Table1-1. Specifications

RATIO (Option 0801

Ratio Measurement Input Configuration:

3 wire; External Reference Low is conlmon with Input Low. External Reference High may be positive or negative with

respect to

Low.

Input

 

Ratio =

 

 

Ext. Ref.

External Reference lnput Resistance: > 107O

DC/DC Measurement Accuracy:

Ext. Ref. Range

! (Ao/oof

Reading r Bo/oof lnput

Range + -

Ext. Ref . Voltage

/

External Reference Ranges: 1 V : 1 O . 1V t o t 1 . 2 V

1 0 V : t l V t o 1 1 2 V

{1 V range should be used for greater accuracy if the Ext- Ref' voltage is between 1 V and 1.2 V.

Input Ranges:

DC: o.1 V to 1000 V

AC: 1 V to 1000 V

xC%of Input Range)

lnput Range

24 hours(23"C I 1"c)

3o days(23"c r 5"c)

9o days(23"c t s"cl

6 months(23oc r s"cl

1 year123oct 5"c)

 

 

1 V

 

 

1 V ,1 0 V

 

100v, 1000v

Ext. Ref.

 

B

c

A

B

c

A

B

c

Range

A

t o v

.003

.001

.oo2

.002 .001 .oo2

.003

.001

.002

1 V

.o03

.002

.002

.002 .002 .002

.005

.003

.002

1 0 v

.oo7 .005 .0m

.002 .003 .002

.005

.003

.002

1 V

. 0 0 8 . 0 1 5 . 0 0 3

.o03 .012 .003

.006 .o12 .003

1 0 v

.o07 .oo5 .002

.002 .003 .002

.006

.o03

.002

1 V

.008 .015 .003

.o03 .012 .OO3

_007 .012

.003

1 0 v

.007

.005

.002

.002 .003 .oo2

.007

.003

.002

1 V

.008

.015

.003

.003 .012 .003

.008

.012

.OO3

1 0 v

.o07

.005

.002

.003 .o03 .002

.008

.003

.o02

1 V

.008

.o15

.o04

.004 .012 .004

. o 0 9 . 0 1 2 . O O 4

Notes:

1.On the lOOOV range, add 0.04 PPM/volt to the % of reading specification'

2.EMF's generated external to the 34904 {lay b9 cgmpensated to achieve the % of range accuracy specified by utilizing the rear panel ThermafAdjust provided'

DC/DC Temperature Coeff icient:

 

 

 

 

 

 

 

 

 

 

! lo/oof Reading + % of Input

Range)/-C

 

 

 

 

 

 

 

 

 

 

 

 

lnput Range

 

 

 

 

Ext. Ref.

 

 

1 V

 

1 V , 1 0 V

 

100v, 1000v

 

Range

 

 

 

 

 

'10

 

r

{0.0003+ 0.0007}

r

(0.0001+ 0.00041

=

(0.0005

 

 

v

 

+ 0.0004)

1 V

 

r

(0.0004+ 0.0021

r

(0.0002

t

(0.0006+ 0.002)

 

 

+ 0.0021

 

 

 

 

DC/DC Measurement Accuracy Example:

 

AC/DC

Measurement Accuracy :

 

 

lnput Voltage = 0.5 V; Input Range= 1 V

 

10 V Ext. Ref. Range: Same as AC Voltage aocuracy

Ext. Ref. Voltage = 0.5 V; Ext. Ref . Range= 1 V

 

1 V Ext. Ref. Range: ^

Same^asAC Voltage accuracv

30 day spec. = t (0.0037oof

Reading + O.O12%of lnput

Range

24 hours 123-C t 1-C)l

Ext. Ref. Range

 

 

 

 

30 davs to 1 year Q3-C x 5-C): Add

o.01 to % of Rang

x 0.003% of Input Rangel

 

 

AC Voltage accuracy

 

 

Ext. Ref . Voltage

 

 

 

 

{SeeTable 1-1 of main 34904 manual.l

 

Ratio Readingshould be 1.00000

 

 

 

 

 

 

 

 

 

0.003% of Reading

=

3 counts

 

AC/DC Temperature Coeff icient:

 

 

O.O12%of Input Range

: 12 counts

 

10V

Ext. Ref.

Range: Same as

AC

Voltage temperature

x o.oos"Ä of Input Range=

6 counts

 

coefficient

 

 

o/o

 

 

 

 

 

 

 

*. 5 V

 

 

 

 

1 V

Ext. Ref. Range: Add

0.001 to

ol

Range in AC Voltage

Total rror tolerance

= 21 counts

 

temperature coeff icient specification

 

Display should be 0.99979 to

1.00021

 

(SeeTable 1-1 of main 3490A manual.)

 

 

 

 

 

 

 

 

tt - t