Model3490AADJUSTMENTPROCEDURESSectionV

a.Removethe toP guard cover.

b.Locate the white/black wire just to the rear of the display cable connector. This wire is labeled "5th Digit Blank" in Figure 5-10. Disconnect this wire from Test Point L and connect to Test Point M. This restoresthe display to five full digits for Sample/Hold measurements, providing greater resolution in order to make the following adjustmentsmore accurately.

c.Set 3490A FTINCTIONto DC, RANGE to IO V. SAMPLE/HOLD to TRACK/HOLD, SAMPLE RATE to HOLD. Short input Higfr to Low.

d.Connect digital voltmeter between A27TP3 and groundtest point on AlAl. Recordvoltagereading.

e.Connectjumper wire betweenTestPoint LTST on A28 and ground.

f.Set SAMPLE/HOLD switch to OFF. then back to

TRACK/HOLD.

g.Adjust A27Rl4, Gain Adjust, for digital voltmeter readingthe sameasthat noted in step d I 0.0005 V.

h.Dsconnect digital voltmeter and remove jumper wire betweenLTST and ground.

5-73.0ffsetAdjustment.

5-74. This procedure adjusts the Offset Amplifier so that there is no turnover error between positive and negativeinputs. A dc standard(-hp 7a0B) is required.

a. With 3490A set to IOVDC RANGE, TRACK/ HOLD, and input shortedas in previousadjustment,set SAMPLE RATE fully clockwise.

b.Adjust A27Rl3, Offset,for 3490A displayof zero t 0.0001v.

c.Dsconnect input short and connect dc standard to input terminals. Set standard output to + 10.0000V. Note 3490A reading.

d.Reversepolarity of input (to - 10.0000V) and note reading- Disregarding polarity display, adjust

A27Rl3 to split the difference between the readingsin

stepc and stepd.

Ii'

 

e.Continue reversinginput polarity and adjusting A27Rl3 until there is no numericaldifferencebetween readingswith positive and negativeinputs.

$75. DielectricAbsorptionAdjustment.

i. Replacetop guard cover and securewith only two

5:76.A dc standard(hp- 7a0B)is requiredfor this

procedure, which adjuststhe dielectric absorption com-

screws.

pensation.

sth DtGtT

BLANK

A27TP3

A27R14

A 2 7 R 1 3

Figure5-10. Locationof Sample/HoldAdlustments.

5-17