Function Mode and Adjustment
6
6.Referring to the setting procedure for Function mode, set "F7".
7.To perform the adjustment, press the Start key.
8.Rotate VR61 on PU2 so that the reading voltage on the multi- meter becomes DC21.5±0.2V.
9.Cancel the Function Mode.
10.Check the Projection Lamp voltage in reader mode is DC17.0±0.3V.
VR61
PU2
NOTE
In order to prevent damage of the Projection Lamp be careful of the following items in adjustment.
➜Do not exceed the DC21.7V during Projection Lamp voltage adjustment.
➜Makes full lighting time in F7 mode of the Projection Lamp into less than 30 seconds.
6.8.8 F8: Reset the Setting Data
Function mode setting value and job program are reset to the default value.
1.Sets the function mode “F8”.
2.Press the Mode Memory Key.
3.A display window blinks and setting data is cleared.
101