F-30

TROUBLESHOOTING & REPAIR

F-30

SNUBBER RESISTOR TEST (cont.)

FIGURE F.9 - SNUBBER RESISTOR TEST POINTS

W R 401/

1/

9

12

4/

402/ R W

403

8

 

 

5

404

L8604-1 SWITCH

Section TOC

Master TOC

5.Test for 25 ohms resistance from lead 401 to termi- nal 12 on Switch Board.

a.If 25 ohms is measured, Resistor R4 is OK.

b.If 30 ohms or more is measured, Resistor R4 is faulty and must be replaced.

c.If 20 ohms or less is measured, Resistor R4 is faulty and must be replaced.

6.Repeat same procedures to test R5, R6, and R7 per Table F.4.

Return to

Return to Section TOC

Return to

Return to Master TOC

TABLE F.4 SNUBBER RESISTORS TEST

Check

 

Test Result

 

Conclusion

 

Next Test Step

 

Repair Action

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lead 401 to

 

25 ohms

 

OK

 

Continue

 

 

 

Terminal 12

 

>30 ohms

 

R4 open

 

 

 

Replace R4

 

 

 

<20 ohms

 

R4 faulty

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lead 402 to

 

25 ohms

 

OK

 

Continue

 

 

 

Terminal 9

 

>30 ohms

 

R5 open

 

 

 

Replace R5

 

 

 

<20 ohms

 

R5 faulty

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lead 403 to

 

25 ohms

 

OK

 

Continue

 

 

 

Terminal 12

 

>30 ohms

 

R6 open

 

 

 

Replace R6

 

 

 

<20 ohms

 

R6 faulty

 

 

 

 

 

Lead 404 to

 

25 ohms

 

OK

 

Continue

 

 

 

Terminal 9

 

>30 ohms

 

R7 open

 

 

 

Replace R7

 

 

 

<20 ohms

 

R7 faulty

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 