Performance Verification4-4 WFM90D and WFM91D Service ManualTable4- 2: Equipment required for performance verification (Cont.)
Equipment description Equipment exampleMinimum requirements
Function generator Range: 0.1 to 5 Vpp when loaded by 75 ,
10Vpp when unl oaded;
Frequency:50 kHz t o 10 MHz.
TektronixFG 5033Function Generator
Audio generator Amplitude Range: --10dBu to 8 dBu;
Frequency:1 kHz t o 50 kHz
TektronixASG 100 Audio Signal Generator
Voltmeter Range:0 VDC to >100 VDC;
Accuracy:0.1%
TektronixDM 504A3Digital Multimeter
Frequencycounter Range: 10 Hz to 10 MHz;
Accuracy:0.001%
TektronixDC 503A3Universal Counter
Videoampl itude calibrator
(VAC)
Signal: Adjustable square wave 0.0 mVpp to
999.9 mVpp; Resolution: 0.1 mV; Accuracy:
0.05%; Frequency: Approximately 270 Hz
Tektronixpart number 067-0916-003
Peak-to-peakdetect or Input Signal Range: 0.25 Vpp to 1.0 Vpp;
Flatness: 0.2% 50 kHz to 10 MHz;
Input Impedance: 75
Tektronixpart number 015-0408-00 (includes
peak-to-peakdetect orhead, Tektronixpart
number 015-0413-00)3
75termination Two required; feed-through type Tektronixpart number 011-0103-02
75coaxi al cable Twocables required, male to male BNC
connectors
Tektronixpart number 012-0159-00 (42-inch)
50coaxi al cable 1%precision, mal e to male BNC connectors Tektronixpart number 012-0482-00 (36-inch)
Audio cable XLRfemale to XLR male Switchcraft part number SC3XXJ
50-to-75 minimum loss
attenuator
Impedance: 50 -to-75; Tolerance: 0.5 dB;
Frequency:100 M Hz
Tektronixpart number 011-0057-00
X10at tenuator 75Tektronixpart number011-0061-00
Dual input coupler MatchedBNC cabl e-Tf orm aking phase
comparisons between two inputs; Matched
length of the two arms within 0.1 inch
Tektronixpart number 067-0525-02
BNCf emale to BNC female
connector
Tektronixpart number 103-0028-00
2The1410-Series generators with standard SPG and TSG modules can be used, but not all checks and adjustments can be
made. The standardSPG2 and SPG12 m odules will not check lock to changes in sync amplitude, cw lock to changes in
burstampli tude, and frequencylock to burst offset frequency changes. The signal generator can be ordered with one or
bothoptions (AA and AB). The TSG3 and TSG13 are Modulated Staircase Generators with variable APL. The TSG5 and
TSG15are Pulse and Bar Generators with modulated pulse and field square wave signals.
3Requiresa Tektronix TM 500-Series or TM 5000-Series Power Module Mainframe.