Model 9060Z Manual

Caution, risk of danger. See additional information in the manual.

3.1HARDWARE SPECIFICATIONS

3.1.1TRANSMITTER SPECIFICATIONS

 

 

 

2 maximum

 

Number of oxygen probes:

 

 

 

0‐1% to 0‐100%

 

Oxygen range:

 

 

 

±1% of actual measured oxygen value with a repeatability of ±0.5% of

 

Oxygen accuracy:

 

 

 

 

the measured value

 

 

Type K, J, R and S

 

Thermocouple types:

 

 

 

±2°C

 

Temperature accuracy:

 

 

 

0‐20mA or 4‐20mA field selectable

 

Analogue outputs:

 

 

 

 

Active outputs (Do NOT loop power these outputs)

 

 

1000 ohm max

 

Output load:

 

 

 

4

 

Alarm relays:

 

 

 

 

2A/240VAC, 2A/30VDC

 

Alarm relay contacts:

 

(WARNING: Do not use both mains voltage and low voltage connections to adjacent alarm contacts)

 

 

 

100 to 240VAC, ‐6 +10%, 50/60 Hz

 

Mains voltage supply:

 

 

 

Category II (IEC60364‐4‐443)

 

Overvoltage:

 

 

 

5W for controller plus probe power

 

Power:

 

 

 

 

530W max., 25% duty cycle each probe on 240VAC

 

 

 

110W max., 100% duty cycle each probe on 110VAC

 

 

 

2.5A max

 

 

3A, fast blow, 250v, 20x5mm (heater fuses, 2 of)

 

Fuses:

 

 

 

 

1A, slow blow, 250v, microfuse (PCB mtg fuse, 1 of)

 

 

Operating Temperature ‐25°C to 55°C

 

Environmental Rating:

 

 

 

 

Relative Humidity 5% to 95% (non‐condensing)

 

 

2000m maximum

 

Altitude

 

 

 

IP65

 

Degree of Protection:

 

 

 

 

IP54 with internal reference air pump

 

 

260mm (10.2”) wide, 160mm (6.3”) high, 90mm (3.5”) deep

 

Case Size:

 

 

 

 

3 Kg (6.6 lbs.)

 

Weight:

WARNING: All signal level connections onto the transmitter must be treated as safety extra‐low voltage (SELV) as defined in the standard IEC61140. Double insulation must be used when connecting these terminals to systems that might carry high voltage.

3.1.2 STANDARD ‘U’ LENGTHS

9060H

9060UL/UH

250 mm (10”)

500 mm (20”)

350 mm (14”)

750 mm (30”)

500 mm (20”)

1000 mm (40”)

750 mm (30”)

1500 mm (60”)

1000 mm (40”)

 

1500 mm (60”)

2000 mm (80”) Ordering Information