Model9060ZManual
10
Caution,riskofdanger.Seeadditionalinformationinthemanual.
3.1HARDWARESPECIFICATIONS

3.1.1TRANSMITTERSPECIFICATIONS

Numberofoxygenprobes: 2maximum
Oxygenrange:
01%to0100%
Oxygenaccuracy: ±1%ofactualmeasuredoxygenvaluewitharepeatabilityof±0.5%of
themeasuredvalue
Thermocoupletypes: TypeK,J,RandS
Temperatureaccuracy: ±2°C
Analogueoutputs: 020mAor420mAfieldselectable
Activeoutputs(DoNOTlooppowertheseoutputs)
Outputload: 1000ohmmax
Alarmrelays: 4
Alarmrelaycontacts: 2A/240VAC,2A/30VDC
(WARNING:Donotusebothmainsvoltageandlowvoltageconnectionstoadjacentalarmcontacts)
Mainsvoltagesupply: 100to240VAC,‐6+10%,50/60Hz
Overvoltage: CategoryII(IEC603644443)
Power: 5Wforcontrollerplusprobepower
530Wmax.,25%dutycycleeachprobeon240VAC
110Wmax.,100%dutycycleeachprobeon110VAC
2.5Amax
Fuses: 3A,fastblow,250v,20x5mm(heaterfuses,2of)
1A,slowblow,250v,microfuse(PCBmtgfuse,1of)
EnvironmentalRating: OperatingTemperature‐25°Cto55°C
RelativeHumidity5%to95%(noncondensing)
Altitude 2000mmaximum
DegreeofProtection: IP65
IP54withinternalreferenceairpump
CaseSize: 260mm(10.2”)wide,160mm(6.3”)high,90mm(3.5”)deep
Weight: 3Kg(6.6lbs.)
WARNING:Allsignallevelconnectionsontothetransmittermustbetreatedassafetyextralow
voltage(SELV)asdefinedinthestandardIEC61140.Doubleinsulationmustbeusedwhen
connectingtheseterminalstosystemsthatmightcarryhighvoltage.

3.1.2STANDARD‘U’LENGTHS

9060H9060UL/UH
250mm(10”)   500mm(20”)
350mm(14”)   750mm(30”)
500mm(20”)   1000mm(40”)
750mm(30”)   1500mm(60”)
1000mm(40”)
1500mm(60”)
2000mm(80”)
OrderingInformation