14. MAINTENANCE

14.1 TRANSMITTER MAINTENANCE

The 9060Z transmitter has several hardware checking systems that confirm the correct operation of the transmitter and raise an alarm if there is a fault detected. Because of this system, the transmitter only requires periodic maintenance. Teledyne Analytical Instruments recommends an annual check of the transmitter.

The annual check should include the following items –

1.Measure and record the reference voltages, and check that the correct voltages are set in the calibration menu

2.Check that the reference air flow from the port on the bottom of the transmitter case is between 100 and 300cc/m

3.Check for contamination and potential blockage of the reference air filter that is mounted on the bottom of the transmitter case

4.If an external burner interlock has been connected (terminals 10 & 11), check that if one of the wires is disconnected the probe heater is disabled (9060H probes only)

5.Check the integrity of the cable. Particularly check the main power supply cable and the probe cable for any damage.

6.Check the integrity of the earth connections in the transmitter

14.2 CLEANING

The 9060Z transmitter is made in an IP65 / IP54 rated diecast case. It is painted with a tough ripple finish power coating to maintain a clean look. However, if it is necessary to clean the case use either a warm soapy water solution or an isopropyl alcohol.

A build up of dust on or around the transmitter should be removed before the cabinet door is opened. If there is a buildup of dust inside the cabinet check the door o‐ring seal for damage and replace it if necessary.

Take care when wiping the display window to avoid scratching the surface.

If the filters of the probe (if fitted) become blocked, use an ultrasonic cleaner with a few drops of household detergent. If the filters cannot be cleaned this way replace the sintered filters.

The filter in the base of the analyzer may become blocked if the analyzer is operating in a dusty environment. It is available as a field replacement item.

14.3 REPLACEMENT PARTS

The following list identifies field replaceable parts –

Stock number

18020

18021

15039

14105

14093

18064

11040

11041

14123

Description

1732‐1 Main PCB

1732‐2 Display PCB

MP‐24E Reference air pump

Fuse ‐M205 3.15A 250V fuse (pack of 5)

Instruction manual

EEROM 9060x firmware

FIL‐230 Titanium sintered filter, 30um

FIL‐215 Titanium sintered filter, 15um

Filter – analyzer bulkhead

54

Page 54
Image 54
Teledyne 9060Z manual Transmitter Maintenance, Cleaning, Replacement Parts