All of the items in this list are selected as the default setting. Any of these items can be disabled from the common alarm by pressing the ENTER key, and they will then appear on the list in the other alarm relays.

7.2.22 Operation of the Alarm Relays when an Alarm is Accepted

An option is available to in the commissioning menu to change the operation of the alarm relay when an alarm has been accepted. The states for the alarm relay contacts are shown in the following table –

Alarm state

Accepted Relay hold setting‐

Accepted Relay hold setting‐

 

Enabled

Disabled

No alarm condition

Closed circuit

Closed circuit

New alarm

Open circuit

Open circuit

All alarms accepted

Open circuit

Closed circuit

Alarms self cleared

Closed circuit

Closed circuit

The default setting is ‘Enabled’ after a cold start.

7.2.23 Selecting the Correct Fuel

The 9060Z transmitter can calculate several measurements that are based on the oxygen reading and the temperature of the process gas. These calculations also require some details about the chemical makeup of the fuel. The additional calculations are:

 

Oxygen deficiency

 

The volume of oxygen required to bring the combustion back to

 

 

 

 

stoichiometry

 

Combustibles

 

 

This theoretical scale is equivalent to a combustibles analyzer

 

 

Carbon dioxide

The calculated maximum level of CO2 in the combustion gas

 

(maximum dry)

 

 

 

 

Combustion

 

Requires a stack thermocouple and so is only available in single probe

 

efficiency

 

mode

Select the required fuel in function 50 by using the OPTION keys and then press the ENTER key. The table of values in the next 6 functions will be updated for that fuel.

7.2.24 Reference Air Pump Options

Normally the reference air is supplied from the transmitter using the internal pump. The default option is 5.00v.

 

Function

Options

57

Reference Air Pump

External or Internal

58

Internal Pump Voltage

Pump voltage setting, between 2.5 and 5.00 in 0.25v steps

59

Reference Air RH%

If external is selected, set the RH level. (5% if instrument

 

 

air is used)