6.3SETUPMODEFUNCTIONS
6.3.1Probe1Offset
EachTeledyneAnalyticalInstrumentsprobehasanoffsetcalibrationvalueprintedonatagthatis
attachedtotheprobewhenitisdispatched.Toachieveaccuratemeasurementtheoffsetvaluemust
beenteredwiththesamepolarityasitisprintedonthislabel.Theoffsetvalueshouldbewithin
±1.0mV
NOTE:Anoffsetof1.0mVwillchangetheoxygenreadingbyapproximately1%oxygenwhenthe
probeisinambientair.However,astheprocessoxygenmeasurementdrops,thisoffsetwillhavea
reducedaffect.Ataprocessgasoxygenconcentrationof2%,the1.0mVoffseterrorwillonlychange
thereadingby0.1%oxygen.
Ifinanydoubtaboutthecorrectoffsetvalue,setitto0.0mV.
Thefunction’Probe2Offset’willonlyappearifthetransmitterhasbeenconfiguredfor2probes.(see
chapter7.2.3NumberofProbes)
6.3.2LowerLineItems
Thisfunctionallowstheoperatortochangetheitemsthatareavailabletobedisplayedonthelower
lineofthetransmitterinRUNmode.Iftheword“Enabled”appearsonthedisplayforaselectedlower
lineoption,themeasurementwillbeavailabletobeshownonthedisplayintheRUNmodeby
scrollingthroughthelistusingtheDISPLAYUPandDISPLAYDOWNkeys.
Alowerlinemeasurementcanbe“Enabled”ordisabledbypressingtheENTERkey.