Model9060ZManual
8
2.INTRODUCTION

2.19060ZTRANSMITTER

TheTeledyneAnalyticalInstruments9060Zoxygentransmitterisdesignedtomeasuretheoxygengas
concentrationincombustionapplicationssuchaspowerstationsandindustrialboilers.The
measurementisusedtocontroltheair/fuelrationintheburnertomaintainefficiencyandsafe
operationconditions.
ThetransmitterisdesignedtobeusedinconjunctionwithaTeledyneAnalyticalInstrumentsprobe
model9060Hor9060UL/UH.Thesetwoprobesuseazirconiasensorthatmustoperateat>650°C.
The9060Hprobehasanintegratedheater,andthe9060UL/UHprobeusestheprocessheattoheat
thesensor.Thereisauniquecableconnectionbetweenthe9060Ztransmitterandthe
9060H/9060UL/UHprobe.
Themodel9060Zoxygentransmitterisbasedonthewellknownmodel9060Htransmitter.Itincludes
anumberofhardwareandsoftwareimprovementssuchasagraphicdisplay,largercharacters,faster
microprocessor,simplifiedsetupmenu,alarmlogging,fasterprobeheatercontrolandmore
calculatedvaluessuchasburnerefficiency.
The9060ZOxygenTransmitterhasavarietyofuserselectablefunctions.Theyaresimpletouse
becauseeachselectionismenudriven.Foroptionsyouarenotsureabout,readthemanualonthat
particulariteminchapter6,SetupMode.
Featuresinclude:‐
Inputs
Twozirconiaoxygenprobes,heatedorunheated
Oxygen
read01%to0100%
Furnace,kilnorfluethermocouple,fieldselectableastypeK,J,RorS
Mainflameestablishedsafetyinterlock(forheatedprobesonly)
Purgepressureorflowswitch
Outputs
Twolinearised420mAor020mADCisolatedoutputs,max.load1000Ω
Theoutputfunctionandtherangearefieldselectable
Commonalarmrelay
Threeotheralarmrelayswithselectablefunctions
DigitalInterface
RS232ortwowireRS485MODBUSforconnectiontoacomputer/DCS/PCLfordiagnosticsofthe
transmitter,probeorcombustionprocess.
Display
Multifontgraphicaldisplay
Largefontcharactersfortheoxygenonthetopline
Selectablelowerlineitemsforthesecondarydisplayfunctions.ieProbetemperature,Oxygensecond
probeAlarmdisplaymodethatshowsthetimeofthealarm,theacceptancetimeandthetimethat
thealarmwasclearedofupto4000alarmevents

2.2SERIES9060HOXYGENPROBES&SENSORS

Pleasenotethatifthisequipmentisnotinstalledandusedinthemannerdescribedinthismanualthen
thesafetyprotectionprovidedbytheequipmentmaybeimpaired.