ImportantNoticeRegarding9060 ProbeOption‐FIL‐3
WARNING:TheonlyidentifiablestandardforflamearrestersforgeneraluseisBritishStandardBS
EN12874:2001.BritishStandardBSEN12874:2001referstoanoperatingenvironmentupto150
DegreesCentigrade.
TheFIL‐3deviceoptionallyfittedto9060HeatedZirconiaProbes(the“Probes"or"Probe")
operateinanenvironmentconsiderablygreaterthan150DegreesCentigrade.
Therefore,weknowofnoAustralian,British,EuropeanorUSAstandardapplicabletoflame
arrestersortheirtestingabove150degreesCentigrade.Consequently,theFIL‐3devicecannotbe
certifiedasasafetydevice.
Theprobeisonlyonprobesduringthestartupprocessesofacombustionappliance.
TheTeledyneAnalyticalInstrumentsBurnerInterlockRelayfacility,whichisastandardpartofthe
TeledyneAnalyticalInstrumentstransmitter,isdesignedtobewiredtothemainsafetyshut‐off
fuelvalvesin���awaythatcanshutdowntheprobeheaterwhenthefuelvalvesareclosed.
TheriskofignitionofflammablegasmixtureatthehotendoftheProbecanonlybeminimizedby
correctuse,maintenanceandoperationoftheFIL‐3device.TheuseroftheFIL‐3deviceis
responsibleforverificationandmaintenanceandcorrectuseandoperationoftheFIL‐3device.
THEUSERAGREESTHATITUSESTHEPROBEANDTHEFIL‐3DEVICEATITSSOLERISK.TELEDYNE
ANALYTICALINSTRUMENTSTOTHEFULLEXTENTPERMITTEDBYLAW,GIVESNOWARRANTIESOR
ASSURANCESANDEXCLUDESALLLIABILITY(INCLUDINGLIABILITYFORNEGLIGENCE)INRELATION
TOTHEPROBEANDTHEFIL‐3DEVICE.
TheusermustensurethatitcorrectlyfollowsallinstructionsinrelationtotheProbeandFIL‐3
device,correctlyunderstandsthespecificationsoftheProbeandFIL‐3deviceandensuresthatthe
ProbeandFIL‐3deviceareregularlyinspectedandmaintained.
FIL‐3equippedProbesshouldbeinspectedatleastonceayearforcorrosionandmorefrequently
ifthereisanyreasontosuspectthatcorrosionmayhaveoccurred.