Model9060ZManual
6


ImportantNoticeRegarding9060 ProbeOption‐FIL3

WARNING:TheonlyidentifiablestandardforflamearrestersforgeneraluseisBritishStandardBS
EN12874:2001.BritishStandardBSEN12874:2001referstoanoperatingenvironmentupto150
DegreesCentigrade.
TheFIL3deviceoptionallyfittedto9060HeatedZirconiaProbes(the“Probes"or"Probe")
operateinanenvironmentconsiderablygreaterthan150DegreesCentigrade.
Therefore,weknowofnoAustralian,British,EuropeanorUSAstandardapplicabletoflame
arrestersortheirtestingabove150degreesCentigrade.Consequently,theFIL3devicecannotbe
certifiedasasafetydevice.
Theprobeisonlyonprobesduringthestartupprocessesofacombustionappliance.
TheTeledyneAnalyticalInstrumentsBurnerInterlockRelayfacility,whichisastandardpartofthe
TeledyneAnalyticalInstrumentstransmitter,isdesignedtobewiredtothemainsafetyshutoff
fuelvalvesin���awaythatcanshutdowntheprobeheaterwhenthefuelvalvesareclosed.
TheriskofignitionofflammablegasmixtureatthehotendoftheProbecanonlybeminimizedby
correctuse,maintenanceandoperationoftheFIL3device.TheuseroftheFIL3deviceis
responsibleforverificationandmaintenanceandcorrectuseandoperationoftheFIL3device.
THEUSERAGREESTHATITUSESTHEPROBEANDTHEFIL3DEVICEATITSSOLERISK.TELEDYNE
ANALYTICALINSTRUMENTSTOTHEFULLEXTENTPERMITTEDBYLAW,GIVESNOWARRANTIESOR
ASSURANCESANDEXCLUDESALLLIABILITY(INCLUDINGLIABILITYFORNEGLIGENCE)INRELATION
TOTHEPROBEANDTHEFIL3DEVICE.
TheusermustensurethatitcorrectlyfollowsallinstructionsinrelationtotheProbeandFIL3
device,correctlyunderstandsthespecificationsoftheProbeandFIL3deviceandensuresthatthe
ProbeandFIL3deviceareregularlyinspectedandmaintained.
FIL3equippedProbesshouldbeinspectedatleastonceayearforcorrosionandmorefrequently
ifthereisanyreasontosuspectthatcorrosionmayhaveoccurred.