APPENDIX 2, PROBE EMF TABLES

ZIRCONIA OXYGEN SENSOR OUTPUT (mV) PROBE TYPE 9060H, SENSOR

 

 

OXYGEN

 

 

 

 

Probe EMF

 

 

OXYGEN

 

 

Probe EMF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

%

 

 

 

@ 720°C (1320°F)

 

 

%

 

 

@ 720°C (1320°F)

 

 

 

20.95

 

 

0

 

 

10

 

 

15.93

 

 

 

 

 

 

 

 

 

20.5

 

 

0.517

 

 

9.5

 

17.033

 

 

 

20

 

 

1.046

 

 

9

 

 

18.196

 

 

19.5

 

 

1.589

 

 

8.5

 

19.426

 

 

 

19

 

 

2.147

 

 

8

 

 

20.73

 

 

18.5

 

 

2.719

 

 

7.5

 

22.12

 

 

 

18

 

 

3.306

 

 

7

 

 

23.607

 

 

17.5

 

 

3.911

 

 

6.5

 

25.204

 

 

 

17

 

 

4.533

 

 

6

 

 

26.93

 

 

16.5

 

 

5.173

 

 

5.5

 

28.808

 

 

 

16

 

 

5.834

 

 

5

 

 

30.867

 

 

15.5

 

 

6.515

 

 

4.5

 

33.145

 

 

 

15

 

 

7.219

 

 

4

 

 

35.695

 

 

14.5

 

 

7.947

 

 

3.5

 

38.59

 

 

 

14

 

 

8.7

 

 

3

 

 

41.94

 

 

13.5

 

 

9.481

 

 

2.5

 

45.913

 

 

 

13

 

 

10.292

 

 

2

 

 

50.797

 

 

12.5

 

 

11.134

 

 

1.5

 

57.135

 

 

 

12

 

 

12.011

 

 

0.9

 

 

66.182

 

 

11.5

 

 

12.925

 

 

0.4

 

82.168

 

 

 

11

 

 

13.881

 

 

0.2

 

 

99.518

 

 

10.5

 

 

14.881

 

 

0.1

 

114.347

 

 

'K' Type TC 29.965 mV @ 720°C (1320°F)

 

 

 

 

 

These tables are based on the Nernst equation: O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EMF =

 

 

 

 

 

.

 

 

Where T is temperature° K

 

 

 

.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 