5.2 OXYGEN DISPLAY UNITS

The measured oxygen content displayed on the top line of the transmitter for probe 1 and on the lower line for probe 2 can be shown in either % or parts per million (ppm). See chapter 6.3.3 Oxygen Display Units for how to change the oxygen display.

The format of the oxygen display changes to maintain the best resolution for the measurement. If ‘Oxygen %’ is selected the transmitter will display oxygen in the following format:

Range

Display format

30.0 to 100.0%

##.#%

1.00 to 29.99%

##.##%

0.100 to 0.999%

#.###%

< 0.100%

scientific notation (#.## x 10 ‐ ## % )

If ‘Oxygen ppm’ is selected the transmitter will show the oxygen in ppm format between 10 and 9,999ppm. Outside of this range the oxygen will be shown in % format described above.

Display format for ‘Oxygen ppm’:

Range

Display format

> 10,000ppm (1%)

% format (see ‘Oxygen %’ section above)

1,000 to 9,999ppm

#,### ppm

100.0 to 999.9ppm

###.# ppm

10.00 to 99.99ppm

##.## ppm

< 10ppm

scientific notation (#.## x 10 ‐ ## % )

5.3 KEYPAD

There are eight keys built into the decal on the outside of the door of the 9060Z transmitter. The key functions are written in BLACK and WHITE to identify the function of the key in either RUN mode or SETUP mode.

Key text

RUN mode (WHITE text)

SETUP mode (BLACK text)

SETUP / RUN

Enter SETUP mode

Return to RUN mode

DISPLAY / FUNCTION UP

Display Next Lower Line Item

Next Function

DISPLAY / FUNCTION

Display Prev. Lower Line Item

Previous Function

DOWN

 

 

ALARM / OPTION UP

Next Alarm

Next / Increment Option

ALARM / OPTION DOWN

Previous Alarm

Previous / Decrement Option

ALARM ACCEPT / ENTER

Acknowledge Displayed Alarm

Save Current Option

GAS 1 PURGE 1 / SENS IMP

Gas 1 / Purge 1 manual activate

Check Probe impedance

GAS 2 PURGE 2 / AUTO CAL

Gas 2 / Purge 2 manual activate

Manually Activate Output Calibration

5.3.1 Keypad in RUN mode

When the transmitter is turned on and has completed its start‐up procedure, it will enter RUN mode and enable the keypad. The key functions in this mode are as follows;