6. SETUP MODE
The SETUP mode is mode is accessed by pressing the SETUP key momentarily so the words ‘Setup Mode’ appear at the bottom of the display. The transmitter will return to the RUN mode when the SETUP key is pressed again or 60 seconds after the last key press.
6.1 FUNCTION SUMMARY TABLE
When the transmitter is in the SETUP mode the SETUP light will be on and the words “Setup Menu” will be shown at the bottom of the display.
The following table shows the SETUP functions:
Menu # | Function name | Range | Default value |
1 | Probe 1 Offset | ‐6.0 to +6.0mV | 0.0mV |
2 | Probe 2 Offset | ‐6.0 to +6.0mV | 0.0mV |
3 | Lower Line Items | See Chapter 6.3.2 |
|
4 | Oxygen Display Units | Oxygen % / Oxygen ppm | Oxygen % |
5 | Damping Factor | No Damping to 10 Samples | 5 Samples |
6 | Process Alarms | Enabled / Disabled | Disabled |
6.2 CHANGING OPTIONS
The purpose of having a menu driven SETUP mode is to allow for configuration of the transmitter using the graphical display and keypad.
Once an option is changed and entered using the ENTER key that value immediately becomes active. It is also stored into the non‐volatile memory within the BBRAM and will be loaded again at device start‐up. To reset the transmitter configuration defaults see chapter 10.2, Cold Start.
To change an option in the SETUP menu system:
1.Select the SETUP mode by pressing the SETUP / RUN key once. The SETUP light will come on and the display will have the format shown below. The operations of the keys are now the operations written in WHITE on the keypad. The menu name is written at the bottom of the display.
2.When the SETUP mode has been selected the required function can be found by using the FUNCTION UP and FUNCTION DOWN keys.
The options available for that function can be seen by using the OPTION UP & OPTION DOWN keys.
3.When the required option is on the display press the ENTER key to save that value.
When finished, press the SETUP / RUN key to return to the RUN mode.