Model 9060Z Manual

7.2.16 Solenoid 1 & 2 Post Freeze

Duration of time the transmitter waits before resuming live readings following a gas calibration or purge event. (See also chapter 7.2.10)

7.2.17 Oxygen Content Calibration Gas 1 & 2

When configured as a gas calibration check solenoid this value specifies the oxygen content of the calibration gas.

7.2.18 Maximum Calibration Gas 1 & 2 Positive / Negative Error

Specifies the maximum positive & negative error levels used during the gas calibration check.

If the difference between the specified oxygen content and the measured process gas exceed these thresholds, a gas calibration error alarm will be triggered immediately.

7.2.19 Process Alarms

The oxygen measurements are continuously monitored by the transmitter for process alarm conditions (Process alarms must be enabled in the Set up menu). The limits and the alarm time delays are configurable for the Very Low Oxygen alarm, Low Oxygen alarm, High Oxygen alarm and the Oxygen Deviation alarm.

7.2.20 Alarm Relay 1, 2 and 3 Function

There are 3 user configurable alarm relays. Any of the 3 relays can be configured to be triggered on the following alarm conditions. In addition, any of the alarm conditions that are disabled from the common alarm relay can also be configured to trigger these 3 relays. Multiple selections can be made.

 

Process Alarms

Oxygen 1 Low

Probe 1 Temperature Low

Oxygen 2 Low **

Probe 2 Temperature Low **

Oxygen 1 Very Low

Calibration 1 In Progress

Oxygen 2 Very Low ** Calibration 2 In Progress **

Oxygen 1 High

Purge 1 In Progress

Oxygen 2 High **

Purge 2 In Progress **

Oxygen Deviation

 

7.2.21 Common Alarm Relay Function

The common alarm relay can be configured to be triggered on any of the following:

Instrument Alarms

Probe 1 Heater Fail

Reference Air Pump Fail

Probe 2 Heater Fail **

Reference Air Pump Overload

Probe 1 High Impedance

Alarm Log Fail

Probe 2 High Impedance **

Output 1 Fail

Probe 1 Thermocouple Open Circuit

Output 2 Fail

Probe 2 Thermocouple Open Circuit **

Heater 1 SSR Fail

Auxiliary Thermocouple Open Circuit

Heater 2 SSR Fail

** Are only available in dual probe mode.

Heater SSR Leakage

Probe 1 Filter Blocked Probe 2 Filter Blocked **

Gas 1 Calibration Error Gas 2 Calibration Error **

39