8 Technical Data
8.1 Product Parameters
OXYGEN
Oxygen | 20.9 - 96% |
Measurement |
|
Range: (from a |
|
concentrator) |
|
Oxygen | ±1.5 % of full scale at constant temperature |
Measurement | and optimal flow* |
Accuracy: |
|
Oxygen | 0.1% Oxygen |
Measurement |
|
Resolution: |
|
See Effects of Low Flow, page 11 of this operating manual for information on the effects of low flow.
FLOW
Flow Measurement | 0 - 10 LPM |
|
Range: |
|
|
Flow Measurement | ±0.2 LPM |
|
Accuracy: |
|
|
Flow Measurement | 0.1 LPM |
|
Resolution: |
|
|
PRESSURE | PSI | KPA |
Pressure | 0.5 - 50 | 3.4 - 344 |
Measurement Range: |
|
|
Pressure | ±0.5% | ±0.5% |
Measurement |
|
|
Accuracy: |
|
|
|
|
|
|
|
|
|
| Technical Data | |
|
|
|
|
|
|
|
|
|
|
Pressure | 0.1 |
|
|
|
|
|
| 0.1 up to 199, 1 |
|
Measurement |
|
|
|
|
|
|
| from 200 to 344 |
|
Resolution: |
|
|
|
|
|
|
|
|
|
Response Time: | ≤ 17 seconds |
|
|
| |||||
< 1 second |
|
|
| ||||||
Operating | 59° F - 104° F (15° C - 40° C) |
| |||||||
Temperature: |
|
|
|
|
|
|
|
|
|
Storage | 5° F - 140° F |
| |||||||
Temperature: |
|
|
|
|
|
|
|
|
|
Pressure: | 800 - 1000 mBars |
| |||||||
Humidity: | 0 - 95% |
| |||||||
Power |
| ||||||||
Requirements: |
|
|
|
|
|
|
|
|
|
Battery Life: | ≥1,100 hours (16,500 read cycles) |
| |||||||
Low Battery | “Low Battery” icon displayed on LCD |
| |||||||
Indication: |
|
|
|
|
|
|
|
|
|
Dimensions: | 3.16” x 5.10” x 1.04” (80.3 mm x 129.5 |
| |||||||
| mm x 26.4 mm) |
| |||||||
Weight: | 0.4 lbs (181 grams) |
| |||||||
CLASSIFICATIONS |
|
|
|
|
|
|
|
|
|
Protection against | Internally powered equipment |
| |||||||
electrical shock: |
|
|
|
|
|
|
|
|
|
Protection against |
|
|
|
|
|
|
|
|
|
water: |
|
|
|
|
|
|
|
|
|
Mode of operation: | Continuous |
|
|
| |||||
Sterilization: | See 7.1 Cleaning, page 14. |
| |||||||
Flammable | Not for use in presence of flammable |
| |||||||
anaesthetic mixture: | anaesthetic mixtures |
| |||||||
Power specification: | 1.8 - 3.2 V |
|
|
|
|
| 32 mW 10mA |
| |
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
15 |