4-2 Storing Units

The units or modes can be selected and stored in the function table. The sequence of displaying the units or modes can be arranged in the function table so as to fit the frequency of use.

Select a unit or mode and arrange the sequence of display as follows:

1Press and hold the SAMPLE key until function table is displayed.

ba5fnc

of the

2 Press the SAMPLE key several times to display

1Unit .

3 Press the PRINT key to enter the unit selection mode.

4Specify a unit or mode in the order to be displayed using the following keys.

SAMPLE

key

To sequentially display the units.

 

 

 

RE-ZERO

key

To specify a unit or mode. The

 

 

 

stabilization

indicator appears

 

 

 

when the displayed unit or mode

 

 

 

is specified.

 

5Press the PRINT key to store the units or modes. The balance displays end and then displays the next menu of the function table.

6Press the CAL key to exit the function table. Then the balance returns to the weighing mode with the unit specified first in step 4.

Notes

When the power is turned on, the scale displays the unit specified first in step 4.

In the weighing mode, to select a unit or mode for weighing, press the MODE key.

11

Image 13
Contents WM+PD4001199 RI SPage Contents RS-232C Serial Interface MaintenanceConnection to Peripheral Equipment External Dimens IonsIntroduction FeaturesAbout This Manual ComplianceUnpacking Unpacking the BalanceInstalling the Balance Before Use PrecautionsDuring Use After Use Power Supply Display symbols Display Symbols and Key Operation Key operation Function TableUnits Weighing UnitsCapacity Sample Storing UnitsRE-ZERO Weighing Basic Operation Gram ModeSelecting the counting mode Counting Mode PCSStoring a sample unit mass Counting operation Counting mode using the Acai functionSelecting the percent mode Percent Mode %Storing the 100% reference mass Reading the percentageGetting Started Statistical Calculation ModeSelecting the unit Selecting the statistical items to outputMode PrintOutputting the statistical results Clearing the statistical data Deleting the latest dataEnabling the statistical calculation mode Statistical Calculation Mode Example of UseReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Operation Response AdjustmentFast MID SlowCalibration Calibration ModeTerms DisplayCAL Calibration Using an External WeightEnd Calibration Test Using an External Weight Glp1 Switches Operation Function Switch and InitializationPermit or Inhibit Initializing the Balance Structure and Sequence of the Function Table Function TableDisplay and Keys ExampleDetails of the Function Table Auto print polarity Data output modeAuto print difference Data outputAK, Error code TimeoutMode output items Factory settingsDescription of the Class Environment, Display Hold function Hold Animal weighing modeCondition Cond CondTrc Zero tracking trcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Description of the Item Data Format Interval output modeStandard format DP Dump print format 5if typeMT format KF format 5if typeNU numerical format CSV format 5if typeData Format Examples Statistical calculation mode apf Capacity indicator apfDescription of the Item Application Function Comparator Function Setting exampleID Number and GLP Report Setting the ID NumberGLP Report Display character setCalibration test report using an external weight Calibration report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT Sub 9-pin assignments RS-232C Serial InterfaceConnection to the AD-8121B Printer Connection to Peripheral EquipmentOFF Connection to a Computer Using Windows Communication Tools WinCTRsCom RsKeyUsing the WinCT software, the balance can do the following Command List CommandsCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Error Codes Asking For Repair Options Page FX-300i SpecificationsFX-120i / 200i External DimensionsFX-1200i / 2000i GLP TERMS/INDEXStabilization indicator Index Keys and symbols ON/OFF key Sample keyCapacity indicator Condition Counting modeDigit 31, 35 NET indicatorFXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Parity Pcs Stability band width Stop bit Stream modePermit PrinterMemo Memo Memo Page Korea Limited German Scales Office