13.RS-232C SERIAL INTERFACE

The balance is a Data Communication Equipment (DCE) device. Connect the balance to a personal computer (DTE) using a straight through cable.

Transmission system :

EIA RS-232C (D-Sub 9-pin, female connector)

 

Transmission form

:

Asynchronous, bi-directional, half duplex

 

Transmission rate

:

20 times/second, 10 times/second or 5 times/second (same as data refresh rate)

Data format

: Baud rate

 

 

: 600, 1200, 2400, 4800, 9600, 19200 bps

 

 

 

Data bits

 

 

: 7 or 8 bits

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parity

:

Even, Odd

(Data bits 7 bits)

 

 

 

 

 

 

 

 

 

 

None

 

 

 

 

 

(Data bits 8 bits)

 

 

 

Stop bit

:

1 bit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Code

 

 

: ASCII

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RS-232C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

-5V to -15V

 

 

 

 

 

 

 

LSB

 

 

 

 

 

 

 

 

 

MSB

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

1

 

2

3

 

 

4

5

6

 

 

 

 

 

 

 

 

0

+5V to +15V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Data bits

 

 

 

 

 

 

 

 

 

 

Stop bit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Start bit

 

 

 

 

 

 

 

 

 

 

 

 

Parity bit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

D-Sub 9-pin assignments

Pin No.

Signal name

Direction

Description

1

-

-

No connection

2

TXD

Output

Transmit data

3

RXD

Input

Receive data

4

-

-

No connection

5

SG

-

Signal ground

6

DSR

Output

Data set ready

7

RTS

Input

Request to send

8

CTS

Output

Clear to send

9

-

-

No connection

Signal names of the balance side are the same as the DTE side with TXD and RXD reversed.

#4-40UNC

Balance side

 

 

Computer side

DCE

 

 

DTE

 

TXD

2

RXD

 

 

 

 

RXD

3

TXD

 

 

 

 

RTS

7

RTS

 

 

 

 

CTS

8

CTS

 

 

 

Approx.5V

DSR

6

DSR

 

 

 

SG

5

SG

 

 

 

 

 

 

Terminals sample

Inside of balance

 

 

47

Image 49
Contents WM+PD4001199 RI SPage Contents RS-232C Serial Interface MaintenanceConnection to Peripheral Equipment External Dimens IonsIntroduction FeaturesAbout This Manual ComplianceUnpacking Unpacking the BalanceInstalling the Balance Before Use PrecautionsDuring Use After Use Power SupplyDisplay symbols Display Symbols and Key OperationKey operation Function TableUnits Weighing UnitsCapacity Sample Storing UnitsRE-ZERO Weighing Basic Operation Gram ModeSelecting the counting mode Counting Mode PCSStoring a sample unit mass Counting operation Counting mode using the Acai functionSelecting the percent mode Percent Mode %Storing the 100% reference mass Reading the percentageGetting Started Statistical Calculation ModeSelecting the unit Selecting the statistical items to outputMode PrintOutputting the statistical results Clearing the statistical data Deleting the latest dataEnabling the statistical calculation mode Statistical Calculation Mode Example of UseReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Operation Response AdjustmentFast MID SlowCalibration Calibration ModeTerms DisplayCAL Calibration Using an External WeightEnd Calibration Test Using an External Weight Glp1 Switches Operation Function Switch and InitializationPermit or Inhibit Initializing the Balance Structure and Sequence of the Function Table Function TableDisplay and Keys ExampleDetails of the Function Table Auto print polarity Data output modeAuto print difference Data outputAK, Error code TimeoutMode output items Factory settingsDescription of the Class Environment, Display Hold function Hold Animal weighing modeCondition Cond CondTrc Zero tracking trcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Description of the Item Data Format Interval output modeStandard format DP Dump print format 5if typeMT format KF format 5if typeNU numerical format CSV format 5if typeData Format Examples Statistical calculation mode apf Capacity indicator apfDescription of the Item Application Function Comparator Function Setting exampleID Number and GLP Report Setting the ID NumberGLP Report Display character setCalibration test report using an external weight Calibration report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT Sub 9-pin assignments RS-232C Serial InterfaceConnection to the AD-8121B Printer Connection to Peripheral EquipmentOFF Connection to a Computer Using Windows Communication Tools WinCTRsCom RsKeyUsing the WinCT software, the balance can do the following Command List CommandsCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Error Codes Asking For Repair Options Page FX-300i SpecificationsFX-120i / 200i External DimensionsFX-1200i / 2000i GLP TERMS/INDEXStabilization indicator Index Keys and symbols ON/OFF key Sample keyCapacity indicator Condition Counting modeDigit 31, 35 NET indicatorFXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Parity Pcs Stability band width Stop bit Stream modePermit PrinterMemo Memo Memo Page Korea Limited German Scales Office