7-2 Calibration Using an External Weight
This function calibrates the balance using an external weight.
Operation
1Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the pan.
2Press and hold the CAL key until displayed, and then release the key.
Calout
is
3 The balance displays
Cal 0 .
If you want to change the calibration weight (a list of usable weights is shown on page 23), press the SAMPLE key and proceed to step 4.
If you use the calibration weight value stored in the balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE | key | To switch the display condition | |||||||||||
|
|
|
|
| to: | “All | of | the | segments | ||||
|
|
|
|
| blinking” | (calibration | weight | ||||||
|
|
|
|
| selection mode) or “The last | ||||||||
|
|
|
|
| two | digits | blinking” | (value | |||||
|
|
|
|
| adjustment mode). |
|
| ||||||
|
|
|
|
|
|
|
| ||||||
| key (To increase the value) |
|
| ||||||||||
|
|
|
|
|
|
|
| ||||||
MODE | key (To decrease the value) |
|
|
|
| ||||||||
|
|
|
|
| To | select | the |
|
| calibration | |||
|
|
|
|
| weight or adjust the value. | ||||||||
|
|
|
|
|
|
|
|
| |||||
key | To | store | the | new | weight | ||||||||
|
|
|
|
| value. Even | if | the AC | ||||||
|
|
|
|
| adapter is removed, the data | ||||||||
|
|
|
|
| is maintained in | ||||||||
|
|
|
|
| memory. |
|
|
|
|
|
| ||
|
|
|
|
| |||||||||
CAL | key | To cancel the operation and | |||||||||||
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
| return to | Cal | 0 | . |
|
|