10.ID NUMBER AND GLP REPORT
The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is maintained in
The GLP output format is selected at “GLP output (info)” of the function table and can be output to a personal computer or printer using the
The GLP output format includes the balance manufacturer, model, serial number, ID number and space for signature for weight data, and the weight used and results for calibration or calibration test data.
When the
The balance can output the following reports for GLP. “Calibration report” of the calibration, using an external weight. “Calibration test report” of the calibration test, using an external weight. “Title block” and “End block” for the weight data.
10-1 Setting the ID Number
1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed.
2 | Press the | SAMPLE | key several times to display id . | |||||||
|
|
|
|
|
|
|
|
| ||
3 | Press the | key. Set the ID number using the following keys. | ||||||||
|
|
|
|
|
|
|
|
| ||
|
| SAMPLE | key | To select the digit to change the value. | ||||||
|
|
|
|
|
|
|
| |||
|
|
| key | To set the character of the digit selected. Refer to the display character | ||||||
|
|
|
|
|
|
|
| set shown below. | ||
|
|
|
|
|
|
| ||||
|
| key | To store the new ID number and display | ap fnc |
|
CAL | key | To cancel the new ID number and display |
4Press the CAL key to return to the weighing mode.
ap fnc
Display character set
10-2 GLP Report
Set the following parameters to output the report.
To print the report, set the “GLP output ( info )” parameter to “1” and use MODE 3 of the
To output the report to a personal computer using the
Note
For operational details about calibration and calibration test, refer to “7. CALIBRATION”.