8-2 Initializing the Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100% reference mass value (percent mode) External calibration weight
Function switch settings
Statistical data
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The balance displays p5 .
3 Press the SAMPLE key to display
Clr
.
4Press the PRINT key.
To cancel this operation, press the CAL key.
5 Press the
6Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.
29