5-5 Statistical Calculation Mode (Example of Use)

Here, as an example of use of the statistical calculation mode, mixing of the multiple formulae such as medicine is described. The mixing process is recorded using the balance and the printer.

In the example, the FX-3000i and the AD-8121B (dump print mode) are connected using the RS-232C serial interface.

5-5-1 Getting Started

Changing the function table

Changes: ●To enable the statistical calculation mode

●To enable “Zero after output”

Enabling the statistical calculation mode

1 Enter the function table menu.

Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.

2 Select the application function.

Press the SAMPLE key several times to display 1ap fnc . Then, press the PRINT key to display

1°apf 0 .

3 Change the application function parameter to “2”.

Press the RE-ZERO key to display apf 2 . Press the PRINT key to confirm the change. After end , 1ba5fnc1 is displayed.

Enabling “Zero after output”

4Select “Zero after output”.

Press the SAMPLE key several times to display

1dout

.

Then, press the PRINT key to display

°prt 0

, and press

the SAMPLE key several times to display

1°ar-d 0

.

5 Enable “Zero after output”.

Press the RE-ZERO key to display ar-d 1 . Then, press the PRINT key to confirm the change. After 1end , 5if is displayed.

Returning to the weighing mode

6 Press the CAL key to return to the weighing mode.

20

Image 22
Contents RI S WM+PD4001199Page Contents Connection to Peripheral Equipment MaintenanceRS-232C Serial Interface External Dimens IonsAbout This Manual FeaturesIntroduction ComplianceUnpacking the Balance UnpackingInstalling the Balance Before Use PrecautionsDuring Use Power Supply After UseKey operation Display Symbols and Key OperationDisplay symbols Function TableWeighing Units UnitsCapacity Sample Storing UnitsRE-ZERO Basic Operation Gram Mode WeighingSelecting the counting mode Counting Mode PCSStoring a sample unit mass Counting mode using the Acai function Counting operationStoring the 100% reference mass Percent Mode %Selecting the percent mode Reading the percentageStatistical Calculation Mode Getting StartedMode Selecting the statistical items to outputSelecting the unit PrintOutputting the statistical results Deleting the latest data Clearing the statistical dataReturning to the weighing mode Statistical Calculation Mode Example of UseEnabling the statistical calculation mode Changing the function tableUsing The Statistical Calculation Mode Fast Response AdjustmentOperation MID SlowTerms Calibration ModeCalibration DisplayCalibration Using an External Weight CALEnd Calibration Test Using an External Weight Glp1 Switches Operation Function Switch and InitializationPermit or Inhibit Initializing the Balance Display and Keys Function TableStructure and Sequence of the Function Table ExampleDetails of the Function Table Auto print difference Data output modeAuto print polarity Data outputMode output items TimeoutAK, Error code Factory settingsCondition Cond Hold function Hold Animal weighing modeDescription of the Class Environment, Display CondDisplay refresh rate 5pd Zero tracking trcTrc Decimal point pntDescription of the Item Data Output Mode Standard format Interval output modeDescription of the Item Data Format DP Dump print format 5if typeNU numerical format KF format 5if typeMT format CSV format 5if typeData Format Examples Statistical calculation mode apf Capacity indicator apfDescription of the Item Application Function Setting example Comparator FunctionGLP Report Setting the ID NumberID Number and GLP Report Display character setCalibration report using an external weight Calibration test report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT RS-232C Serial Interface Sub 9-pin assignmentsConnection to the AD-8121B Printer Connection to Peripheral EquipmentOFF RsCom Using Windows Communication Tools WinCTConnection to a Computer RsKeyUsing the WinCT software, the balance can do the following Commands to query weight data CommandsCommand List Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Error Codes Asking For Repair Options Page Specifications FX-300iFX-120i / 200i External DimensionsFX-1200i / 2000i TERMS/INDEX GLPCapacity indicator Index Keys and symbols ON/OFF key Sample keyStabilization indicator Condition Counting modeFXi-02 FXi-08 FXi-09 FXi-10 FXi-11 NET indicatorDigit 31, 35 GLP output 23, 25, 26, 32Permit Stability band width Stop bit Stream modeParity Pcs PrinterMemo Memo Memo Page German Scales Office Korea Limited