7. CALIBRATION

7-1 Calibration Mode

The FX-i series balance has the following two modes. Calibration using an external weight

Calibration test using an external weight (Calibration test does not perform calibration.)

Terms

The following terms are defined as follows:

External weight

= A weight that you have. Referred to as a calibration weight when used for

 

 

calibration.

Calibration weight

=

A weight used for calibration

Target weight

=

An external weight used for calibration test

Caution

Calibration adjusts the balance for accurate weighing.

Besides periodic calibration and before each use, perform calibration when:

The balance is installed for the first time.

The balance has been moved.

The ambient environment has changed.

Do not allow vibration or drafts to affect the balance during calibration.

To output the data for GLP using the RS-232C serial interface, set “GLP output ( info )” of “Data output ( dout )”. For details, refer to “9. FUNCTION TABLE”.

Calibration test is available only when “GLP output ( info )” of “Data output ( dout )” is set to “1” or “2”.

Caution on using an external weight

The accuracy of an external weight will influence the accuracy of weighing. Select an appropriate weight as listed below:

Model

Usable calibration weight

Adjustable range

FX-120i

100 g, 50 g

 

FX-200i

200 g, 100 g

-0.050 g to +0.050 g

FX-300i

300 g, 200 g, 100 g

 

FX-1200i

1000 g, 500 g,

 

FX-2000i

2000 g, 1000 g

-0.50 g to +0.50 g

FX-3000i

3000 g, 2000 g, 1000 g

 

The calibration weight in bold type: factory setting

The calibration weight value can be adjusted within the range above.

Display

￿This indicator means “the balance is measuring calibration data”.

Do not allow vibration or drafts to affect the balance while this indicator is displayed.

23

Image 25
Contents WM+PD4001199 RI SPage Contents RS-232C Serial Interface MaintenanceConnection to Peripheral Equipment External Dimens IonsIntroduction FeaturesAbout This Manual ComplianceUnpacking Unpacking the BalanceInstalling the Balance Before Use PrecautionsDuring Use After Use Power SupplyDisplay symbols Display Symbols and Key OperationKey operation Function TableUnits Weighing UnitsCapacity Sample Storing UnitsRE-ZERO Weighing Basic Operation Gram ModeSelecting the counting mode Counting Mode PCSStoring a sample unit mass Counting operation Counting mode using the Acai functionSelecting the percent mode Percent Mode %Storing the 100% reference mass Reading the percentageGetting Started Statistical Calculation ModeSelecting the unit Selecting the statistical items to outputMode PrintOutputting the statistical results Clearing the statistical data Deleting the latest dataEnabling the statistical calculation mode Statistical Calculation Mode Example of UseReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Operation Response AdjustmentFast MID SlowCalibration Calibration ModeTerms DisplayCAL Calibration Using an External WeightEnd Calibration Test Using an External Weight Glp1 Switches Operation Function Switch and InitializationPermit or Inhibit Initializing the Balance Structure and Sequence of the Function Table Function TableDisplay and Keys ExampleDetails of the Function Table Auto print polarity Data output modeAuto print difference Data outputAK, Error code TimeoutMode output items Factory settingsDescription of the Class Environment, Display Hold function Hold Animal weighing modeCondition Cond CondTrc Zero tracking trcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Description of the Item Data Format Interval output modeStandard format DP Dump print format 5if typeMT format KF format 5if typeNU numerical format CSV format 5if typeData Format Examples Statistical calculation mode apf Capacity indicator apfDescription of the Item Application Function Comparator Function Setting exampleID Number and GLP Report Setting the ID NumberGLP Report Display character setCalibration test report using an external weight Calibration report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT Sub 9-pin assignments RS-232C Serial InterfaceConnection to the AD-8121B Printer Connection to Peripheral EquipmentOFF Connection to a Computer Using Windows Communication Tools WinCTRsCom RsKeyUsing the WinCT software, the balance can do the following Command List CommandsCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Error Codes Asking For Repair Options Page FX-300i SpecificationsFX-120i / 200i External DimensionsFX-1200i / 2000i GLP TERMS/INDEXStabilization indicator Index Keys and symbols ON/OFF key Sample keyCapacity indicator Condition Counting modeDigit 31, 35 NET indicatorFXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Parity Pcs Stability band width Stop bit Stream modePermit PrinterMemo Memo Memo Page Korea Limited German Scales Office