9-9 Comparator Function
The results of the comparison are indicated by HI OK LO on the display.
Operating conditions: No comparison
| Comparison when the weight data is stable or overloaded, excluding “near zero” |
| Comparison when the weight data is stable or overloaded, including “near zero” |
| Continuous comparison, excluding “near zero” |
| Continuous comparison, including “near zero” |
To compare, use: | Upper limit value and lower limit value |
Input method: | Digital input |
Note
"Near zero" means that the weight value is within ±10 digits of the minimum weighing value. For example, using a
For the description of “Comparator (Cp fnc)”, refer to
Setting example
(Continuous comparison, excluding “near zero”, upper limit and lower limit values, digital input) Selecting a comparator mode
1 Press and hold the SAMPLE key until ba5fnc
2Press the SAMPLE key several times to display
3 Press the PRINT key.
of the function table is displayed.
Cp fnc .
4Press the
5 Press the PRINT key to store the selected mode.
Entering the upper limit value
Cp 3 .
6 With Cp Hi displayed, press the PRINT key. The current setting of the upper limit value is displayed with all the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 7.
When the current setting is to be changed, press the
SAMPLE | key | To select the digit to change the value. | |||
|
|
|
|
| |
| key | To change the value of the digit selected. | |||
|
|
|
| ||
MODE | key | To switch the polarity. | |||
|
|
| |||
key | To store the new setting and go to step 7. | ||||
|
|
| |||
CAL | key | To cancel the new setting and go to step 7. |
Entering the lower limit value
7 With Cp lo displayed, press the PRINT key. The current setting of the lower limit value is displayed with all the digits blinking.
When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 8.
When the current setting is to be changed, press the
SAMPLE | key | To select the digit to change the value. | |||
|
|
|
|
| |
| key | To change the value of the digit selected. | |||
|
|
|
| ||
MODE | key | To switch the polarity. | |||
|
|
| |||
key | To store the new setting and go to step 8. | ||||
|
|
| |||
CAL | key | To cancel the new setting and go to step 8. |
8 Press the CAL key to exit the comparator function and return to the weighing mode.
41