A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Comparator Function, Setting example

Page 43

9-9 Comparator Function

The results of the comparison are indicated by HI OK LO on the display.

Operating conditions: ￿ No comparison

 

￿ Comparison when the weight data is stable or overloaded, excluding “near zero”

 

￿ Comparison when the weight data is stable or overloaded, including “near zero”

 

￿ Continuous comparison, excluding “near zero”

 

￿ Continuous comparison, including “near zero”

To compare, use:

￿ Upper limit value and lower limit value

Input method:

￿ Digital input

Note

"Near zero" means that the weight value is within ±10 digits of the minimum weighing value. For example, using a FX-3000i in gram mode, the range of ±0.10 g is “near zero”.

For the description of “Comparator (Cp fnc)”, refer to “9-3 Details of the Function Table”.

Setting example

(Continuous comparison, excluding “near zero”, upper limit and lower limit values, digital input) Selecting a comparator mode

1 Press and hold the SAMPLE key until ba5fnc

2Press the SAMPLE key several times to display

3 Press the PRINT key.

of the function table is displayed.

Cp fnc .

4Press the RE-ZERO key several times to display

5 Press the PRINT key to store the selected mode.

Entering the upper limit value

Cp 3 .

6 With Cp Hi displayed, press the PRINT key. The current setting of the upper limit value is displayed with all the digits blinking.

￿When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 7.

￿When the current setting is to be changed, press the RE-ZERO key. Change the setting using the following keys.

SAMPLE

key

To select the digit to change the value.

 

 

 

 

 

RE-ZERO

key

To change the value of the digit selected.

 

 

 

 

MODE

key

To switch the polarity.

 

 

 

PRINT

key

To store the new setting and go to step 7.

 

 

 

CAL

key

To cancel the new setting and go to step 7.

Entering the lower limit value

7 With Cp lo displayed, press the PRINT key. The current setting of the lower limit value is displayed with all the digits blinking.

￿When the current setting is not to be changed, press the PRINT or CAL key to proceed to step 8.

￿When the current setting is to be changed, press the RE-ZERO key. Change the setting using the following keys.

SAMPLE

key

To select the digit to change the value.

 

 

 

 

 

RE-ZERO

key

To change the value of the digit selected.

 

 

 

 

MODE

key

To switch the polarity.

 

 

 

PRINT

key

To store the new setting and go to step 8.

 

 

 

CAL

key

To cancel the new setting and go to step 8.

8 Press the CAL key to exit the comparator function and return to the weighing mode.

41

Image 43
Contents WM+PD4001199 RI SPage Contents External Dimens Ions MaintenanceRS-232C Serial Interface Connection to Peripheral EquipmentCompliance FeaturesIntroduction About This ManualUnpacking Unpacking the BalanceInstalling the Balance Before Use PrecautionsDuring Use After Use Power SupplyFunction Table Display Symbols and Key OperationDisplay symbols Key operationUnits Weighing UnitsCapacity Sample Storing UnitsRE-ZERO Weighing Basic Operation Gram ModeSelecting the counting mode Counting Mode PCSStoring a sample unit mass Counting operation Counting mode using the Acai functionReading the percentage Percent Mode %Selecting the percent mode Storing the 100% reference massGetting Started Statistical Calculation ModePrint Selecting the statistical items to outputSelecting the unit ModeOutputting the statistical results Clearing the statistical data Deleting the latest dataChanging the function table Statistical Calculation Mode Example of UseEnabling the statistical calculation mode Returning to the weighing modeUsing The Statistical Calculation Mode MID Slow Response AdjustmentOperation FastDisplay Calibration ModeCalibration TermsCAL Calibration Using an External WeightEnd Calibration Test Using an External Weight Glp1 Switches Operation Function Switch and InitializationPermit or Inhibit Initializing the Balance Example Function TableStructure and Sequence of the Function Table Display and KeysDetails of the Function Table Data output Data output modeAuto print polarity Auto print differenceFactory settings TimeoutAK, Error code Mode output itemsCond Hold function Hold Animal weighing modeDescription of the Class Environment, Display Condition CondDecimal point pnt Zero tracking trcTrc Display refresh rate 5pdDescription of the Item Data Output Mode DP Dump print format 5if type Interval output modeDescription of the Item Data Format Standard formatCSV format 5if type KF format 5if typeMT format NU numerical formatData Format Examples Statistical calculation mode apf Capacity indicator apfDescription of the Item Application Function Comparator Function Setting exampleDisplay character set Setting the ID NumberID Number and GLP Report GLP ReportCalibration test report using an external weight Calibration report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT Sub 9-pin assignments RS-232C Serial InterfaceConnection to the AD-8121B Printer Connection to Peripheral EquipmentOFF RsKey Using Windows Communication Tools WinCTConnection to a Computer RsComUsing the WinCT software, the balance can do the following Commands to control the balance CommandsCommand List Commands to query weight dataAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Error Codes Asking For Repair Options Page FX-300i SpecificationsFX-120i / 200i External DimensionsFX-1200i / 2000i GLP TERMS/INDEXCondition Counting mode Index Keys and symbols ON/OFF key Sample keyStabilization indicator Capacity indicatorGLP output 23, 25, 26, 32 NET indicatorDigit 31, 35 FXi-02 FXi-08 FXi-09 FXi-10 FXi-11Printer Stability band width Stop bit Stream modeParity Pcs PermitMemo Memo Memo Page Korea Limited German Scales Office