A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i instruction manual Troubleshooting

Page 57

17. TROUBLESHOOTING

17-1 Checking the Balance Performance and Environment

The balance is a precision instrument. When the operating environment or the operating method is inadequate, correct weighing can not be performed. Place a sample on the pan and remove it, and repeat this several times. If the balance seems to have a problem with repeatability or to perform improperly, check as described below. If improper performance persists after checking, contact the local A&D dealer for repair.

Checking that the balance performs properly

Check the balance repeatability using an external weight. Be sure to place the weight in the center of the weighing pan.

Check the balance repeatability, linearity and calibrated value using external weights with a known value.

Checking that the operating environment or weighing method is proper

Operating environment

Is the weighing table solid enough (especially for the FX-120i /200i /300i, the balance with a minimum weighing value of 0.001 g)?

Is the balance level? Refer to “3-1 Before Use”.

Is the operating environment free from vibration and drafts? For the FX-120i /200i /300i, has the breeze break been installed?

Is there a strong electrical or magnetic noise source such as a motor near the balance?

Weighing method

Is the weighing pan installed correctly?

Is the RE-ZERO key pressed before placing a sample on the weighing pan? Is the sample placed in the center of the weighing pan?

Has the balance been calibrated using an external weight?

Has the balance been warmed up for 30 minutes before weighing?

Sample and container

Has the sample absorbed or lost moisture due to the ambient conditions such as temperature and humidity?

Has the temperature of the container been allowed to equalize to the ambient temperature? Refer to “3-2 During Use”.

Is the sample charged with static electricity? Refer to “3-2 During Use”.

The FX-120i /200i /300i is prone to be charged with static electricity when the relative humidity is low. Is the sample of magnetic material such as iron? Caution is required for weighing magnetic materials. Refer to “3-2 During Use”

55

Image 57
Contents WM+PD4001199 RI SPage Contents RS-232C Serial Interface MaintenanceConnection to Peripheral Equipment External Dimens IonsIntroduction FeaturesAbout This Manual ComplianceUnpacking Unpacking the BalanceInstalling the Balance Precautions Before UseDuring Use After Use Power SupplyDisplay symbols Display Symbols and Key OperationKey operation Function TableUnits Weighing UnitsCapacity Storing Units SampleRE-ZERO Weighing Basic Operation Gram ModeCounting Mode PCS Selecting the counting modeStoring a sample unit mass Counting operation Counting mode using the Acai functionSelecting the percent mode Percent Mode %Storing the 100% reference mass Reading the percentageGetting Started Statistical Calculation ModeSelecting the unit Selecting the statistical items to outputMode PrintOutputting the statistical results Clearing the statistical data Deleting the latest dataEnabling the statistical calculation mode Statistical Calculation Mode Example of UseReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Operation Response AdjustmentFast MID SlowCalibration Calibration ModeTerms DisplayCAL Calibration Using an External WeightEnd Calibration Test Using an External Weight Glp1 Function Switch and Initialization Switches OperationPermit or Inhibit Initializing the Balance Structure and Sequence of the Function Table Function TableDisplay and Keys ExampleDetails of the Function Table Auto print polarity Data output modeAuto print difference Data outputAK, Error code TimeoutMode output items Factory settingsDescription of the Class Environment, Display Hold function Hold Animal weighing modeCondition Cond CondTrc Zero tracking trcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Description of the Item Data Format Interval output modeStandard format DP Dump print format 5if typeMT format KF format 5if typeNU numerical format CSV format 5if typeData Format Examples Capacity indicator apf Statistical calculation mode apfDescription of the Item Application Function Comparator Function Setting exampleID Number and GLP Report Setting the ID NumberGLP Report Display character setCalibration test report using an external weight Calibration report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT Sub 9-pin assignments RS-232C Serial InterfaceConnection to Peripheral Equipment Connection to the AD-8121B PrinterOFF Connection to a Computer Using Windows Communication Tools WinCTRsCom RsKeyUsing the WinCT software, the balance can do the following Command List CommandsCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Troubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Error Codes Asking For Repair Options Page FX-300i SpecificationsExternal Dimensions FX-120i / 200iFX-1200i / 2000i GLP TERMS/INDEXStabilization indicator Index Keys and symbols ON/OFF key Sample keyCapacity indicator Condition Counting modeDigit 31, 35 NET indicatorFXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Parity Pcs Stability band width Stop bit Stream modePermit PrinterMemo Memo Memo Page Korea Limited German Scales Office