12. PROGRAMMABLE-UNIT

This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary coefficient set in the function table and displays the result.

The coefficient must be within the range between the minimum and maximum shown below. If the coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.

Model

Minimum coefficient

Maximum coefficient

FX-120i /200i /300i

0.000001

1000

FX-1200i /2000i /3000i

100

 

Operation

1 Press and hold the SAMPLE key until

ba5fnc

of the function table is displayed.

2

Press the

SAMPLE

key several times to display

mltj

.

 

 

 

3

Press the

PRINT

key. The balance enters the mode to confirm or set the coefficient.

Confirming the coefficient

4The current coefficient is displayed with the first digit blinking.

￿ When it is not to be changed, press the CAL key and proceed to step 6.

￿ When it is to be changed, press the RE-ZERO key and proceed to step 5.

Setting the coefficient

5 Set the coefficient using the following keys.

SAMPLE

key

To select a digit to change the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

value. The selected digit blinks.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RE-ZERO

key

To change the value.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODE

key

To change the decimal point

 

 

 

 

 

position.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Each time the switch is pressed,

 

 

 

 

 

the decimal

 

point position

 

 

 

 

 

changes as follows:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PRINT

key

To store the new setting,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

display

end

and go to step 6.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL

key

To cancel the new setting and

 

 

 

 

 

go to step 6.

 

 

 

 

 

 

 

 

 

 

 

 

 

Quitting the operation

6The balance displays Unit . Press the CAL key to exit the programmable-unit function and return to the weighing mode.

Using the function

Press the MODE key to select the programmable-unit (no display on the unit section). Perform weighing as described in “5-1 Basic Operation (Gram Mode)”. After weighing, the balance displays the result (weight data in grams x coefficient).

46

Image 48
Contents RI S WM+PD4001199Page Contents Maintenance RS-232C Serial InterfaceConnection to Peripheral Equipment External Dimens IonsFeatures IntroductionAbout This Manual ComplianceUnpacking the Balance UnpackingInstalling the Balance Precautions Before UseDuring Use Power Supply After UseDisplay Symbols and Key Operation Display symbolsKey operation Function TableWeighing Units UnitsCapacity Storing Units SampleRE-ZERO Basic Operation Gram Mode WeighingCounting Mode PCS Selecting the counting modeStoring a sample unit mass Counting mode using the Acai function Counting operationPercent Mode % Selecting the percent modeStoring the 100% reference mass Reading the percentageStatistical Calculation Mode Getting StartedSelecting the statistical items to output Selecting the unitMode PrintOutputting the statistical results Deleting the latest data Clearing the statistical dataStatistical Calculation Mode Example of Use Enabling the statistical calculation modeReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Response Adjustment OperationFast MID SlowCalibration Mode CalibrationTerms DisplayCalibration Using an External Weight CALEnd Calibration Test Using an External Weight Glp1 Function Switch and Initialization Switches OperationPermit or Inhibit Initializing the Balance Function Table Structure and Sequence of the Function TableDisplay and Keys ExampleDetails of the Function Table Data output mode Auto print polarityAuto print difference Data outputTimeout AK, Error codeMode output items Factory settingsHold function Hold Animal weighing mode Description of the Class Environment, DisplayCondition Cond CondZero tracking trc TrcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Interval output mode Description of the Item Data FormatStandard format DP Dump print format 5if typeKF format 5if type MT formatNU numerical format CSV format 5if typeData Format Examples Capacity indicator apf Statistical calculation mode apfDescription of the Item Application Function Setting example Comparator FunctionSetting the ID Number ID Number and GLP ReportGLP Report Display character setCalibration report using an external weight Calibration test report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT RS-232C Serial Interface Sub 9-pin assignmentsConnection to Peripheral Equipment Connection to the AD-8121B PrinterOFF Using Windows Communication Tools WinCT Connection to a ComputerRsCom RsKeyUsing the WinCT software, the balance can do the following Commands Command ListCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Troubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Error Codes Asking For Repair Options Page Specifications FX-300iExternal Dimensions FX-120i / 200iFX-1200i / 2000i TERMS/INDEX GLPIndex Keys and symbols ON/OFF key Sample key Stabilization indicatorCapacity indicator Condition Counting modeNET indicator Digit 31, 35FXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Stability band width Stop bit Stream mode Parity PcsPermit PrinterMemo Memo Memo Page German Scales Office Korea Limited