12. PROGRAMMABLE-UNIT
This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary coefficient set in the function table and displays the result.
The coefficient must be within the range between the minimum and maximum shown below. If the coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.
Model | Minimum coefficient | Maximum coefficient | |
0.000001 | 1000 | ||
100 | |||
|
Operation
1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed.
2 | Press the | SAMPLE | key several times to display | mltj | . | |
|
|
| ||||
3 | Press the | key. The balance enters the mode to confirm or set the coefficient. |
Confirming the coefficient
4The current coefficient is displayed with the first digit blinking.
When it is not to be changed, press the CAL key and proceed to step 6.
When it is to be changed, press the
Setting the coefficient
5 Set the coefficient using the following keys.
SAMPLE | key | To select a digit to change the |
|
|
|
|
|
|
| |||||||||||||
| ||||||||||||||||||||||
|
|
|
|
|
|
| ||||||||||||||||
|
|
|
|
| value. The selected digit blinks. | |||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| key | To change the value. | ||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
MODE | key | To change the decimal point | ||||||||||||||||||||
|
|
|
|
| position. |
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
| Each time the switch is pressed, | |||||||||||||||||
|
|
|
|
| the decimal |
| point position | |||||||||||||||
|
|
|
|
| changes as follows: | |||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
key | To store the new setting, | |||||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| display | end | and go to step 6. | |||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
CAL | key | To cancel the new setting and | ||||||||||||||||||||
|
|
|
|
| go to step 6. |
|
|
|
|
|
|
|
|
|
|
|
|
|
Quitting the operation
6The balance displays Unit . Press the CAL key to exit the
Using the function
Press the MODE key to select the