Display | Error code | Description |
|
|
|
| EC, E00 | Communications error |
|
| A protocol error occurred in communications. |
|
| Confirm the format, baud rate and parity. |
|
|
|
| EC, E01 | Undefined command error |
|
| An undefined command was received. |
|
| Confirm the command. |
|
|
|
| EC, E02 | Not ready |
|
| A received command can not be processed. |
|
| e.g. The balance received a Q command, but not in the |
|
| weighing mode. |
|
| e.g. The balance received a Q command while |
|
| processing a |
|
| Adjust the delay time to transmit a command. |
|
|
|
| EC, E03 | Timeout error |
|
| If the timeout parameter is set to |
|
| did not receive the next character of a command within |
|
| the time limit of one second. |
|
| Confirm the communication. |
|
|
|
| EC, E04 | Excess characters error |
|
| The balance received excessive characters in a |
|
| command. |
|
| Confirm the command. |
|
|
|
| EC, E06 | Format error |
|
| A command includes incorrect data. |
|
| e.g. The data is numerically incorrect. |
|
| Confirm the command. |
|
|
|
| EC, E07 | Parameter setting error |
|
| The received data exceeds the range that the balance |
|
| can accept. |
|
| Confirm the parameter range of the command. |
|
|
|
Other errors |
| If the errors described above can not be released or |
|
| other errors are displayed, contact the local A&D dealer. |
|
|
|
17-3 Asking For Repair
If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe the following when transporting the balance.
Use the original packing material.
Remove the weighing pan from the main unit
57