21. TERMS/INDEX

 

 

Terms

 

Stable value

The weight data when the stabilization indicator appears.

Environment

Ambient conditions such as vibration, drafts, temperature, static electricity and

 

magnetic fields which affect the weighing operation.

Calibration

Adjustment of the balance so that it can weigh accurately.

Output

To output the weight data using the RS-232C serial interface.

Zero point

A weighing reference point or the zero display. Usually refers to the value

 

displayed when nothing is on the weighing pan.

Digit

Unit of digital resolution. Used for the balance, a unit of minimum weighing value.

Tare

To cancel the weight of a container which is not included in the weight data.

Mode

Balance operational function.

Re-zero

To set the display to zero.

GLP

Good Laboratory Practice.

Repeatability

Variation in measured values obtained when the same weight is placed and

 

removed repetitively. Usually expressed as a standard deviation.

 

e.g. Standard deviation=1 digit: This means that measured values fall within ±1

 

digit in the frequency of about 68%.

Stabilization time Time required after a sample being placed, until the stabilization indicator illuminates and the weight data is displayed.

Sensitivity drift An affect that a change in temperature causes to the weight data. Expressed as temperature coefficient.

e.g. Temperature coefficient = 2 ppm/°C : If a load is 300 g and the temperature changes by 10°C, the value displayed changes by the following value. 0.0002%/°C x 10°C x 300 g = 6 mg

62

Image 64
Contents RI S WM+PD4001199Page Contents Maintenance RS-232C Serial InterfaceConnection to Peripheral Equipment External Dimens IonsFeatures IntroductionAbout This Manual ComplianceUnpacking the Balance UnpackingInstalling the Balance Before Use PrecautionsDuring Use Power Supply After UseDisplay Symbols and Key Operation Display symbolsKey operation Function TableWeighing Units UnitsCapacity Sample Storing UnitsRE-ZERO Basic Operation Gram Mode WeighingSelecting the counting mode Counting Mode PCSStoring a sample unit mass Counting mode using the Acai function Counting operationPercent Mode % Selecting the percent modeStoring the 100% reference mass Reading the percentageStatistical Calculation Mode Getting StartedSelecting the statistical items to output Selecting the unitMode PrintOutputting the statistical results Deleting the latest data Clearing the statistical dataStatistical Calculation Mode Example of Use Enabling the statistical calculation modeReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Response Adjustment OperationFast MID SlowCalibration Mode CalibrationTerms DisplayCalibration Using an External Weight CALEnd Calibration Test Using an External Weight Glp1 Switches Operation Function Switch and InitializationPermit or Inhibit Initializing the Balance Function Table Structure and Sequence of the Function TableDisplay and Keys ExampleDetails of the Function Table Data output mode Auto print polarityAuto print difference Data outputTimeout AK, Error codeMode output items Factory settingsHold function Hold Animal weighing mode Description of the Class Environment, DisplayCondition Cond CondZero tracking trc TrcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Interval output mode Description of the Item Data FormatStandard format DP Dump print format 5if typeKF format 5if type MT formatNU numerical format CSV format 5if typeData Format Examples Statistical calculation mode apf Capacity indicator apfDescription of the Item Application Function Setting example Comparator FunctionSetting the ID Number ID Number and GLP ReportGLP Report Display character setCalibration report using an external weight Calibration test report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT RS-232C Serial Interface Sub 9-pin assignmentsConnection to the AD-8121B Printer Connection to Peripheral EquipmentOFF Using Windows Communication Tools WinCT Connection to a ComputerRsCom RsKeyUsing the WinCT software, the balance can do the following Commands Command ListCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Error Codes Asking For Repair Options Page Specifications FX-300iFX-120i / 200i External DimensionsFX-1200i / 2000i TERMS/INDEX GLPIndex Keys and symbols ON/OFF key Sample key Stabilization indicatorCapacity indicator Condition Counting modeNET indicator Digit 31, 35FXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Stability band width Stop bit Stream mode Parity PcsPermit PrinterMemo Memo Memo Page German Scales Office Korea Limited