6. RESPONSE ADJUSTMENT

This function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and vibration at the place where the balance is installed.

The function has three stages as follows and can be changed by simple key operation.

Indicator

Description

FAST

Fast response, but prone to drafts and vibration.

Good for target weighing.

 

MID.

 

SLOW

Slow response, but strong against drafts and vibration

Good for weighing which requires a stable display.

 

Operation

1 Press and hold the MODE key until RESPONSE is displayed. And then, release the key.

2 Press the MODE key to select a rate of the response adjustment. Either FAST, MID. or SLOW can be selected.

3 After a few seconds of inactivity, or when the PRINT

key is pressed, the balance displays eend . Then, it returns to the weighing mode and displays the updated response indicator. The response indicator remains displayed for a while.

Note

Setting the response rate automatically changes the “Condition (Cond)” and “Display refresh rate (5pd)” parameters of “Environment, Display (ba5fnc)” in the function table, as shown below:

Indicator

Cond (Condition)

5pd (Display refresh rate)

FAST

0

2

MID.

1

0

SLOW

2

0

When the balance is to be used with other setting combinations, set each parameter in the function table.

22

Image 24
Contents RI S WM+PD4001199Page Contents Maintenance RS-232C Serial InterfaceConnection to Peripheral Equipment External Dimens IonsFeatures IntroductionAbout This Manual ComplianceUnpacking the Balance UnpackingInstalling the Balance Precautions Before UseDuring Use Power Supply After UseDisplay Symbols and Key Operation Display symbolsKey operation Function TableWeighing Units UnitsCapacity Storing Units SampleRE-ZERO Basic Operation Gram Mode WeighingCounting Mode PCS Selecting the counting modeStoring a sample unit mass Counting mode using the Acai function Counting operationPercent Mode % Selecting the percent modeStoring the 100% reference mass Reading the percentageStatistical Calculation Mode Getting StartedSelecting the statistical items to output Selecting the unitMode PrintOutputting the statistical results Deleting the latest data Clearing the statistical dataStatistical Calculation Mode Example of Use Enabling the statistical calculation modeReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Response Adjustment OperationFast MID SlowCalibration Mode CalibrationTerms DisplayCalibration Using an External Weight CALEnd Calibration Test Using an External Weight Glp1 Function Switch and Initialization Switches OperationPermit or Inhibit Initializing the Balance Function Table Structure and Sequence of the Function TableDisplay and Keys ExampleDetails of the Function Table Data output mode Auto print polarityAuto print difference Data outputTimeout AK, Error codeMode output items Factory settingsHold function Hold Animal weighing mode Description of the Class Environment, DisplayCondition Cond CondZero tracking trc TrcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Interval output mode Description of the Item Data FormatStandard format DP Dump print format 5if typeKF format 5if type MT formatNU numerical format CSV format 5if typeData Format Examples Capacity indicator apf Statistical calculation mode apfDescription of the Item Application Function Setting example Comparator FunctionSetting the ID Number ID Number and GLP ReportGLP Report Display character setCalibration report using an external weight Calibration test report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT RS-232C Serial Interface Sub 9-pin assignmentsConnection to Peripheral Equipment Connection to the AD-8121B PrinterOFF Using Windows Communication Tools WinCT Connection to a ComputerRsCom RsKeyUsing the WinCT software, the balance can do the following Commands Command ListCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Troubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Error Codes Asking For Repair Options Page Specifications FX-300iExternal Dimensions FX-120i / 200iFX-1200i / 2000i TERMS/INDEX GLPIndex Keys and symbols ON/OFF key Sample key Stabilization indicatorCapacity indicator Condition Counting modeNET indicator Digit 31, 35FXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Stability band width Stop bit Stream mode Parity PcsPermit PrinterMemo Memo Memo Page German Scales Office Korea Limited