A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Precautions, Before Use, During Use

Page 8

3. PRECAUTIONS

To get the optimum performance from the balance and acquire accurate weight data, note the following:

3-1 Before Use

Install the balance in an environment where the temperature and humidity are not excessive. The best operating temperature is about 20°C / 68°F at about 50% relative humidity.

Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or air conditioners.

Install the balance where it is free of dust.

Install the balance away from equipment which produces magnetic fields.

Install the balance in a stable place avoiding vibration and shock. Corners of rooms on the first floor are best, as they are less prone to vibration.

The weighing table should be solid and free from vibration, drafts and as level as possible.

Level the balance by adjusting the leveling feet and confirm it using the bubble spirit level.

Ensure a stable power source when using the AC adapter.

Plug in the AC adapter and warm up the balance for at least 30 minutes.

Calibrate the balance before use or after having moved it to another location.

Caution

Do not install the balance where flammable or corrosive gas is present.

3-2 During Use

Discharge static electricity from the material to be weighed (hereinafter referred to as the sample). When a sample could have a static charge, the weight data is influenced. Ground the balance and try the following:

Eliminate static electricity using the optional AD-1683 DC static eliminator.

Try to keep the ambient humidity above 45%RH.

Use a metal shield case for a charged sample. Wipe charged plastic samples with a moistened cloth.

6

Image 8
Contents RI S WM+PD4001199Page Contents Maintenance RS-232C Serial InterfaceConnection to Peripheral Equipment External Dimens Ions Features Introduction About This Manual ComplianceUnpacking the Balance UnpackingInstalling the Balance During Use PrecautionsBefore Use Power Supply After UseDisplay Symbols and Key Operation Display symbolsKey operation Function TableWeighing Units UnitsCapacity RE-ZERO Storing UnitsSample Basic Operation Gram Mode WeighingStoring a sample unit mass Counting Mode PCSSelecting the counting mode Counting mode using the Acai function Counting operationPercent Mode % Selecting the percent modeStoring the 100% reference mass Reading the percentageStatistical Calculation Mode Getting StartedSelecting the statistical items to output Selecting the unitMode PrintOutputting the statistical results Deleting the latest data Clearing the statistical dataStatistical Calculation Mode Example of Use Enabling the statistical calculation modeReturning to the weighing mode Changing the function tableUsing The Statistical Calculation Mode Response Adjustment OperationFast MID SlowCalibration Mode CalibrationTerms DisplayCalibration Using an External Weight CALEnd Calibration Test Using an External Weight Glp1 Permit or Inhibit Function Switch and InitializationSwitches Operation Initializing the Balance Function Table Structure and Sequence of the Function TableDisplay and Keys ExampleDetails of the Function Table Data output mode Auto print polarityAuto print difference Data outputTimeout AK, Error codeMode output items Factory settingsHold function Hold Animal weighing mode Description of the Class Environment, DisplayCondition Cond CondZero tracking trc TrcDisplay refresh rate 5pd Decimal point pntDescription of the Item Data Output Mode Interval output mode Description of the Item Data FormatStandard format DP Dump print format 5if typeKF format 5if type MT formatNU numerical format CSV format 5if typeData Format Examples Description of the Item Application Function Capacity indicator apfStatistical calculation mode apf Setting example Comparator FunctionSetting the ID Number ID Number and GLP ReportGLP Report Display character setCalibration report using an external weight Calibration test report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT RS-232C Serial Interface Sub 9-pin assignmentsOFF Connection to Peripheral EquipmentConnection to the AD-8121B Printer Using Windows Communication Tools WinCT Connection to a ComputerRsCom RsKeyUsing the WinCT software, the balance can do the following Commands Command ListCommands to query weight data Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Checking that the balance performs properly TroubleshootingChecking the Balance Performance and Environment Error Codes Asking For Repair Options Page Specifications FX-300iFX-1200i / 2000i External DimensionsFX-120i / 200i TERMS/INDEX GLPIndex Keys and symbols ON/OFF key Sample key Stabilization indicatorCapacity indicator Condition Counting modeNET indicator Digit 31, 35FXi-02 FXi-08 FXi-09 FXi-10 FXi-11 GLP output 23, 25, 26, 32Stability band width Stop bit Stream mode Parity PcsPermit PrinterMemo Memo Memo Page German Scales Office Korea Limited