5-3 Percent Mode (%)
This is the mode to display the weight value in percentage compared to a 100% reference mass and
is used for target weighing or checking the sample variable.
Selecting the percent mode
1 Press the MODE key to select % (percent mode).
Storing the 100% reference mass
2 Press the SAMPLE key to enter the 100% reference mass storing mode.
Even in the storing mode, pressing the MODE key will switch to the next mode.
3 Place a container on the weighing pan, if necessary. Press the
balance displays
100 0 % .
4Place the sample to be set as the 100% reference mass on the pan or in the container.
5Press the PRINT key to store the reference mass. The balance displays 10000. % . (The decimal point position depends on the reference value. The reference mass stored, even if the AC adapter is removed, is maintained in
Note
If the balance judges that the mass of the sample is too light to be used as a reference, it displays lo . Do not use the sample.
6 Remove the sample.
Reading the percentage
7Place a sample to be compared to the reference mass on the pan. The displayed percentage is based on the 100% reference mass.
While the stabilization indicator is on, pressing the PRINT key will output the weight value, using the
Note
Peripheral equipment, that is sold separately , such as a printer or a personal computer is required.
15