A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Description of the Item Data Format

Page 39

Interval output mode

The weight data is periodically output.

When the PRINT key is pressed, the balance starts to output the weight data at a preset interval time. When the PRINT key is pressed again, the balance stops outputting the weight data.

Required setting

dout

prt

6

Interval output mode

 

dout

int

 

Interval time

Example

“For outputting the weight data periodically."

Caution

The balance may not transmit the data completely at the specified interval times and baud rate. Set the baud rate higher.

9-6 Description of the Item “Data Format”

A&D standard format

5if type 0

This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is used, set the printer to MODE 1 or 2.

￿This format consists of fifteen characters excluding the terminator.

￿A header of two characters indicates the balance condition.

￿The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.

￿The unit, consisting of three characters, follows the data.

DP (Dump print) format

5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B is used, set the printer to MODE 3.

￿This format consists of sixteen characters excluding the terminator.

￿A header of two characters indicates the balance condition. No overload header is used.

￿The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero or overloaded.

￿The unit, consisting of three characters, follows the data.

37

Image 39
Contents WM+PD4001199 RI SPage Contents External Dimens Ions MaintenanceRS-232C Serial Interface Connection to Peripheral EquipmentCompliance FeaturesIntroduction About This ManualUnpacking Unpacking the BalanceInstalling the Balance Precautions Before UseDuring Use After Use Power SupplyFunction Table Display Symbols and Key OperationDisplay symbols Key operationUnits Weighing UnitsCapacity Storing Units SampleRE-ZERO Weighing Basic Operation Gram ModeCounting Mode PCS Selecting the counting modeStoring a sample unit mass Counting operation Counting mode using the Acai functionReading the percentage Percent Mode %Selecting the percent mode Storing the 100% reference massGetting Started Statistical Calculation ModePrint Selecting the statistical items to outputSelecting the unit ModeOutputting the statistical results Clearing the statistical data Deleting the latest dataChanging the function table Statistical Calculation Mode Example of UseEnabling the statistical calculation mode Returning to the weighing modeUsing The Statistical Calculation Mode MID Slow Response AdjustmentOperation FastDisplay Calibration ModeCalibration TermsCAL Calibration Using an External WeightEnd Calibration Test Using an External Weight Glp1 Function Switch and Initialization Switches OperationPermit or Inhibit Initializing the Balance Example Function TableStructure and Sequence of the Function Table Display and KeysDetails of the Function Table Data output Data output modeAuto print polarity Auto print differenceFactory settings TimeoutAK, Error code Mode output itemsCond Hold function Hold Animal weighing modeDescription of the Class Environment, Display Condition CondDecimal point pnt Zero tracking trcTrc Display refresh rate 5pdDescription of the Item Data Output Mode DP Dump print format 5if type Interval output modeDescription of the Item Data Format Standard formatCSV format 5if type KF format 5if typeMT format NU numerical formatData Format Examples Capacity indicator apf Statistical calculation mode apfDescription of the Item Application Function Comparator Function Setting exampleDisplay character set Setting the ID NumberID Number and GLP Report GLP ReportCalibration test report using an external weight Calibration report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT Sub 9-pin assignments RS-232C Serial InterfaceConnection to Peripheral Equipment Connection to the AD-8121B PrinterOFF RsKey Using Windows Communication Tools WinCTConnection to a Computer RsComUsing the WinCT software, the balance can do the following Commands to control the balance CommandsCommand List Commands to query weight dataAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Troubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Error Codes Asking For Repair Options Page FX-300i SpecificationsExternal Dimensions FX-120i / 200iFX-1200i / 2000i GLP TERMS/INDEXCondition Counting mode Index Keys and symbols ON/OFF key Sample keyStabilization indicator Capacity indicatorGLP output 23, 25, 26, 32 NET indicatorDigit 31, 35 FXi-02 FXi-08 FXi-09 FXi-10 FXi-11Printer Stability band width Stop bit Stream modeParity Pcs PermitMemo Memo Memo Page Korea Limited German Scales Office