A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Acknowledge Code and Error Codes

Page 54

15-2 Acknowledge Code and Error Codes

When the “AK, Error code (erCd)” parameter of “Serial interface ( 5if )” is set to “1”, the balance outputs <AK> code or an error code for each command as follows:

<AK> (06h) Acknowledge in ASCII code.

When the balance receives a command to request data and can not process it, the balance transmits an error code (EC, Exx).

When the balance receives a command to request data and can process it, the balance outputs the data.

When the balance receives a command to control the balance and can not process it, the balance transmits an error code (EC, Exx).

When the balance receives a command to control the balance and can process it, the balance transmits the acknowledge code.

Among commands to control the balance, the following transmit the acknowledge code both when the balance receives the command and when the balance has accomplished the command. If the command can not be processed properly, the balance transmits an error code (EC, Exx). This error can be released using the CAL command.

CAL command (Calibration command)

P command (Display ON/OFF command)

ON command (Display ON command)

R command (RE-ZERO command)

R command example

52

Image 54
Contents RI S WM+PD4001199Page Contents Connection to Peripheral Equipment MaintenanceRS-232C Serial Interface External Dimens IonsAbout This Manual FeaturesIntroduction ComplianceUnpacking the Balance UnpackingInstalling the Balance Precautions Before UseDuring Use Power Supply After UseKey operation Display Symbols and Key OperationDisplay symbols Function TableWeighing Units UnitsCapacity Storing Units SampleRE-ZERO Basic Operation Gram Mode WeighingCounting Mode PCS Selecting the counting modeStoring a sample unit mass Counting mode using the Acai function Counting operationStoring the 100% reference mass Percent Mode %Selecting the percent mode Reading the percentageStatistical Calculation Mode Getting StartedMode Selecting the statistical items to outputSelecting the unit PrintOutputting the statistical results Deleting the latest data Clearing the statistical dataReturning to the weighing mode Statistical Calculation Mode Example of UseEnabling the statistical calculation mode Changing the function tableUsing The Statistical Calculation Mode Fast Response AdjustmentOperation MID SlowTerms Calibration ModeCalibration DisplayCalibration Using an External Weight CALEnd Calibration Test Using an External Weight Glp1 Function Switch and Initialization Switches OperationPermit or Inhibit Initializing the Balance Display and Keys Function TableStructure and Sequence of the Function Table ExampleDetails of the Function Table Auto print difference Data output modeAuto print polarity Data outputMode output items TimeoutAK, Error code Factory settingsCondition Cond Hold function Hold Animal weighing modeDescription of the Class Environment, Display CondDisplay refresh rate 5pd Zero tracking trcTrc Decimal point pntDescription of the Item Data Output Mode Standard format Interval output modeDescription of the Item Data Format DP Dump print format 5if typeNU numerical format KF format 5if typeMT format CSV format 5if typeData Format Examples Capacity indicator apf Statistical calculation mode apfDescription of the Item Application Function Setting example Comparator FunctionGLP Report Setting the ID NumberID Number and GLP Report Display character setCalibration report using an external weight Calibration test report using an external weightTitle block and end block Underhook PROGRAMMABLE-UNIT RS-232C Serial Interface Sub 9-pin assignmentsConnection to Peripheral Equipment Connection to the AD-8121B PrinterOFF RsCom Using Windows Communication Tools WinCTConnection to a Computer RsKeyUsing the WinCT software, the balance can do the following Commands to query weight data CommandsCommand List Commands to control the balanceAcknowledge Code and Error Codes Settings Related to RS-232C Maintenance Troubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Error Codes Asking For Repair Options Page Specifications FX-300iExternal Dimensions FX-120i / 200iFX-1200i / 2000i TERMS/INDEX GLPCapacity indicator Index Keys and symbols ON/OFF key Sample keyStabilization indicator Condition Counting modeFXi-02 FXi-08 FXi-09 FXi-10 FXi-11 NET indicatorDigit 31, 35 GLP output 23, 25, 26, 32Permit Stability band width Stop bit Stream modeParity Pcs PrinterMemo Memo Memo Page German Scales Office Korea Limited