Fluke specifications Ruska 7350 Block Diagram

Page 18

RUSKA 7350

Users Manual

Another important component on the Microprocessor Board is the lithium battery, which continuously updates the RUSKA 7350’s date and time, even when the unit is powered down.

The Sensors, Front Panel, the pump expansion and compression limit switches communicate with this board through cables. The Processor Board also supports the RS-232 serial interface, which allows the user’s computer to communicate with the RUSKA 7350.

TEST PORT

AIR

Pneumatic Section

PUMP

SENSORS

GEAR MOTOR

AIR REGULATOR

Electronic Section

MOTOR

76XX INTERFACE

MICROPROCESSOR

DISPLAY/

CONTROLLER

BOARD

BOARD

KEY PAD

 

 

 

PANEL

 

DC POWER

BACK PLANE

IEEE CARD

 

SUPPLY

BOARD

 

 

 

AC POWER

OPTION

 

 

 

BOARD

 

POWER SUPPLY

gld01.eps

Figure 2-1. RUSKA 7350 Block Diagram

2-2

Image 18
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii Introduction How to Contact FlukeGeneral Information General InformationSafety Information Symbols Used in this Manual FeaturesEase of Operation Power-On Self TestAutomatic Zero Adjust Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Electronics Section Power SupplyTheory of Operation Back Plane BoardRuska 7350 Block Diagram IEEE-488 Interface Front PanelMotor Controller Pneumatic Pump Pneumatic Module76xx Interface Board GearmotorSystem Transducer Electronics RegulatorDome Regulator Test PortRuska Installation Unpacking the RuskaPowering Up the Ruska Pneumatic ConnectionsDrive Air Supply Port Numeric Keypad Local OperationLocal Operation Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Mode of Operation Tare ModeSelecting Language Controlling Pressure Setting the Pressure Set-PointMain Menu Triple Range OptionalVent Step/JogMenu Setup MenuLow Limit Menu Setup LimitsHigh Limit Auto VentControl Band Slew Rate AccessMenu Setup User Slew LimitReady Tolerance Step SizeBar Graph Maximum Gas Head Pressure CorrectionChanging the Number of Decimals Menu Setup UnitsPressure Filter Key Click13. Menu Setup Units Edit Name Menu Gpib Address Menu Setup RemoteMenu Setup System ProtocolDate/Time ResetCalibration Password Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMode of Entry MaxTimeEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Capabilities Remote OperationRemote Operation IEEE-488Remote/Local Operation RS-232Device Messages ConfigurationScpi Command Format Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Ruska 7350’s Full Scale Rating MaintenancePreventive Maintenance Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingCleaning Ruska 7350 Pump LubricationFuse Preparation for Storage and Shipment Disconnection InstructionsPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range