Theory of Operation 2
Pneumatic Module
Electronics Regulator
The electronics regulator supplies a 0 – 100 psi air signal for the dome regulator. The electronics regulator is driven by a 0 – 10 VDC signal from the interface board.
Dome Regulator
The dome regulators, which are
System Transducer
The system transducer accurately measures (accuracy is specified as purchased, 0.01% minimum) the controlled pressure. This transducer should be calibrated at a regular interval to maintain its stated accuracy.
Test Port
The device under test is connected to this port. This port is a 1/8 NPT female.
Drive Air-Supply Port
The supply air is connected to the
Gas Supply Port
The clean, dry gas supply is connected to this port. The gas pressure should be 10% above the full scale pressure of the controller. This port is a 1/8 NPT female connection.