Fluke 7350 Electronics Regulator, Dome Regulator, System Transducer, Test Port, Gas Supply Port

Page 21

Theory of Operation 2

Pneumatic Module

Electronics Regulator

The electronics regulator supplies a 0 – 100 psi air signal for the dome regulator. The electronics regulator is driven by a 0 – 10 VDC signal from the interface board.

Dome Regulator

The dome regulators, which are self-relieving, “rough” in the commanded pressure. The precise pressure is then achieved by the pump after closing the system valve.

System Transducer

The system transducer accurately measures (accuracy is specified as purchased, 0.01% minimum) the controlled pressure. This transducer should be calibrated at a regular interval to maintain its stated accuracy.

Test Port

The device under test is connected to this port. This port is a 1/8 NPT female.

Drive Air-Supply Port

The supply air is connected to the drive-air supply port. The supply air should be 100 – 110 psi (7 – 7.5 bar). The pneumatic side is protected by a relief valve set at 120 psi. (8.3 bar). This port is a 1/4 NPT female connection.

Gas Supply Port

The clean, dry gas supply is connected to this port. The gas pressure should be 10% above the full scale pressure of the controller. This port is a 1/8 NPT female connection.

2-5

Image 21
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeIntroduction General InformationSafety Information Features Symbols Used in this ManualAutomatic Zero Adjust Power-On Self TestEase of Operation Easily ProgrammableRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Theory of Operation Power SupplyElectronics Section Back Plane BoardRuska 7350 Block Diagram IEEE-488 Interface Front PanelMotor Controller 76xx Interface Board Pneumatic ModulePneumatic Pump GearmotorDome Regulator Electronics RegulatorSystem Transducer Test PortRuska Unpacking the Ruska InstallationPowering Up the Ruska Pneumatic ConnectionsDrive Air Supply Port Local Operation Local OperationNumeric Keypad Function KeysMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Selecting Mode of Operation Tare ModeSelecting Language Main Menu Setting the Pressure Set-PointControlling Pressure Triple Range OptionalStep/Jog VentMenu Menu SetupHigh Limit Menu Setup LimitsLow Limit Auto VentControl Band Menu Setup User AccessSlew Rate Slew LimitBar Graph Maximum Step SizeReady Tolerance Gas Head Pressure CorrectionPressure Filter Menu Setup UnitsChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup System Menu Setup RemoteGpib Address ProtocolCalibration Password ResetDate/Time Menu CalibratePreparing to Program Menu Program Storing a Sequence in MemoryMode of Entry MaxTimeEntering a New Program 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska Remote Operation Remote OperationCapabilities IEEE-488RS-232 Remote/Local OperationScpi Command Format ConfigurationDevice Messages Scpi Response FormatScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Preventive Maintenance MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsCleaning Ruska 7350 Pump LubricationFuse Preparation for Storage and Shipment Disconnection InstructionsPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of SpecificationsSpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range