Fluke 7350 specifications Calibration Instructions Single Sensor, Preparation

Page 65

Maintenance 6

Calibration

Calibration Instructions — Single Sensor

To calibrate the RUSKA 7350, the user simply connects a calibration standard to the Calibrator’s Test port then follows the 4-step calibration procedure on the Callibrator’s display. The calibration procedures is a menu driven procedure that steps the user through the pressures that are required to calibrate each pressure range that is included in the Calibrator. The user generates the defined pressures utilizing an adequate standard to calibration the RUSKA 7350. The pressure steps will vary based on the pressure range and number of sub ranges that are included with the RUSKA 7350. Typically, for a single range instrument, the operator would be prompted to generate Atmosphere, 50%, 100%, and 50% of the Calibrator’s full scale operating range of a sensor (See Chapter 6, Observing the RUSKA 7350’s Full scale Rating). No disassembly is required and there are no potentiometers to tune.

Note

The uncertainty of the final calibration must include the uncertainty of the standard being used.

Preparation

Verify that the calibration standard is connected to the TEST PORT.

Verify that the RUSKA 7350 has been at stable operating temperature for at least two hours

Verify that the RUSKA 7350 is in Measure mode (See Chapter 4, Local Operation).

Verify and set the head pressure to zero in Menu ⎢Setup ⎢User.

(Triple range only) Verify the range to be calibrated is selected. To change press the Mode ⎢Range [F5].

If desired, change the RUSKA 7350’s units of measure (See Chapter 4) to match those of the calibration standard.

To go to the Calibration screen, select MENU/CAL.

To begin the calibration process, press MENU ⎢CALIBRATE ⎢CALIBRATE. If the calibration access code is enables, enter it at the prompt. The Calibration, Step 1, screen will appear.

Note

To exit the calibration procedure before the calibration coefficients have been changed, press CANCEL any time during the procedure.

Step 1

1.To begin Step 1 of the calibration process, select Step 1. Enter the actual pressure applied and press ok.

2.Wait until the zero procedure completes. When the 7350 completes Step 1, the Calibration, Step 2, screen will appear.

Step 2

1.To begin Step 2, use the calibration standard to Apply the mid-point pressure requested by the RUSKA 7350. As pressure is to be applied into the Test Port, the Measured pressure on the RUSKA 7350’s screen will change accordingly.

2.When the Measured pressure stabilizes, use the RUSKA 7350’s numeric keypad and OK to enter the actual pressure applied by the calibration standard. Do not enter the Measured pressure reported by the RUSKA 7350. If necessary, use the CLEAR key to correct a mistake in the edit field. If the actual pressure applied is acceptable, the Calibration, Step 3, screen will appear.

6-3

Image 65
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeIntroduction General InformationSafety Information Features Symbols Used in this ManualAutomatic Zero Adjust Power-On Self TestEase of Operation Easily ProgrammableRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Theory of Operation Power SupplyElectronics Section Back Plane BoardRuska 7350 Block Diagram Motor Controller IEEE-488 InterfaceFront Panel 76xx Interface Board Pneumatic ModulePneumatic Pump GearmotorDome Regulator Electronics RegulatorSystem Transducer Test PortRuska Unpacking the Ruska InstallationDrive Air Supply Port Powering Up the RuskaPneumatic Connections Local Operation Local OperationNumeric Keypad Function KeysMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Selecting Language Selecting Mode of OperationTare Mode Main Menu Setting the Pressure Set-PointControlling Pressure Triple Range OptionalStep/Jog VentMenu Menu SetupHigh Limit Menu Setup LimitsLow Limit Auto VentControl Band Menu Setup User AccessSlew Rate Slew LimitBar Graph Maximum Step SizeReady Tolerance Gas Head Pressure CorrectionPressure Filter Menu Setup UnitsChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup System Menu Setup RemoteGpib Address ProtocolCalibration Password ResetDate/Time Menu CalibratePreparing to Program Menu Program Storing a Sequence in MemoryEntering a New Program Mode of EntryMaxTime 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska Remote Operation Remote OperationCapabilities IEEE-488RS-232 Remote/Local OperationScpi Command Format ConfigurationDevice Messages Scpi Response FormatScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Preventive Maintenance MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsFuse CleaningRuska 7350 Pump Lubrication Packing Instructions Preparation for Storage and ShipmentDisconnection Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of SpecificationsSpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range