Fluke 7350 specifications Date/Time, Reset, Menu Calibrate, Calibration Password

Page 40

RUSKA 7350

Users Manual

Date/Time

The Calibrator’s system clock is continuously updated. If the date or time requires editing, the following applies;

1.The date and time are set from the Setup System Menu. From the Main Menu (press PREVIOUS until the Main Menu appears), press Menu [F6] Setup [F2] System [F5].

2.To set the system date, turn the rotary knob until Date is highlighted. Use the numeric keypad to enter the current four-digit year, month, and day (yyyymmdd). All digits must be entered. Press ENTER to accept.

3.To set the system time, turn the rotary knob until Time is highlighted. Use the numeric keypad to enter the current hour, minute, and second (hhmmss). All digits must be entered. Press ENTER to accept.

Reset

The Reset [F1] command is used to re-boot the Calibrator.

To reset the system, from the Main Menu (press PREVIOUS until the Main Menu appears), press Menu [F6] Setup [F2] System [F5] Reset [F1].

Menu Calibrate

The Menu Calibrate command is used to perform the calibrations on all of the sensors utilized by the Calibrator. The RUSKA 7350 can be completely calibrated using the front panel interface. It does not require any external computer or software in order to perform a successful calibration. It also has the capability of being calibrated remotely through either the RS232 or IEEE-488 interface. The RUSKA 7350 will guide the operator through a menu driven calibration procedure identifying the pressures that should be supplied by a standard. Following the calibration procedure, the RUSKA 7350 calculates and saves its own coefficients. The calibration coefficients are date stamped so the user can easily identify the last time the instrument went through a full calibration, when the coefficients were last edited or when the Calibrator was re-zeroed.

The calibration section in the RUSKA 7350 can be password protected to prevent any unauthorized entry into the calibration section of the system. Please refer to Chapter 6, Maintenance, of the manual for more detailed information regarding the calibration of the RUSKA 7350.

Calibration Password

The calibration password allows the user to protect access to Calibrator calibration constants and the calibration procedure. If the calibration password is set to any number other than zero, it is required before the user is allowed to calibrate the Calibrator or manually change the calibration constants.

Note

It is recommended that the access password be recorded and filed in a secure location.

1.The calibration password is set from the Calibration Menu. From the Main Menu (press PREVIOUS until the Main Menu appears), press Menu [F6] Calibrate [F3] Menu Calibrate Edit.

2.Press the Access [F4] key. If there is a password in the system other than 0, then you must enter the current valid password into the RUSKA 7350 before it will allow you to modify the password.

4-16

Image 40
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii How to Contact Fluke General InformationIntroduction General InformationSafety Information Symbols Used in this Manual FeaturesPower-On Self Test Automatic Zero AdjustEase of Operation Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Power Supply Theory of OperationElectronics Section Back Plane BoardRuska 7350 Block Diagram Front Panel IEEE-488 InterfaceMotor Controller Pneumatic Module 76xx Interface BoardPneumatic Pump GearmotorElectronics Regulator Dome RegulatorSystem Transducer Test PortRuska Installation Unpacking the RuskaPneumatic Connections Powering Up the RuskaDrive Air Supply Port Local Operation Local OperationNumeric Keypad Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Tare Mode Selecting Mode of OperationSelecting Language Setting the Pressure Set-Point Main MenuControlling Pressure Triple Range OptionalVent Step/JogMenu Setup MenuMenu Setup Limits High LimitLow Limit Auto VentControl Band Access Menu Setup UserSlew Rate Slew LimitStep Size Bar Graph MaximumReady Tolerance Gas Head Pressure CorrectionMenu Setup Units Pressure FilterChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup Remote Menu Setup SystemGpib Address ProtocolReset Calibration PasswordDate/Time Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMaxTime Mode of EntryEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Remote Operation Remote OperationCapabilities IEEE-488Remote/Local Operation RS-232Configuration Scpi Command FormatDevice Messages Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Maintenance Preventive MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingRuska 7350 Pump Lubrication CleaningFuse Disconnection Instructions Preparation for Storage and ShipmentPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range