Fluke 7350 specifications Changing the Configuration Stored with a Program, Menu Program Edit Menu

Page 46

RUSKA 7350

Users Manual

2.Use the rotary knob to highlight the name of the program.

3.Press Edit [F4]. The program editing screen will appear as a table showing all of the steps.

gld22.bmp

Figure 4-20. Menu Program Edit - Menu

4.The rotary knob can be used to move through the program and highlight the value you wish to edit. Use the keyboard to enter a new value. This value will show up in the scratch pad located at the bottom of the window. Press Enter to accept the edited value.

5.To add a step to the program, first move to the step after the new step. For example, to insert a step between steps 3 and 4, move to step 4. Press Insert [F4].

6.To delete a step in the program, first move to the step to be deleted, then press Delete [F5].

7.When all changes have been made, press PREVIOUS to return to the Named Programs screen. The Calibrator automatically saves the changes that were made in the program.

Changing the Configuration Stored with a Program

1.Programs are configured from the Program Menu. From the Main Menu (press PREVIOUS until the Main Menu appears), press Menu [F6] Program [F4].

2.Use the rotary knob to highlight the name of the program.

3.Press Config [F3]. The configuration screen will appear.

4.Press Recall [F2]. The Calibrator will be set to the configuration stored with the program.

4-22

Image 46
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii Introduction How to Contact FlukeGeneral Information General InformationSafety Information Symbols Used in this Manual FeaturesEase of Operation Power-On Self TestAutomatic Zero Adjust Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Electronics Section Power SupplyTheory of Operation Back Plane BoardRuska 7350 Block Diagram Front Panel IEEE-488 InterfaceMotor Controller Pneumatic Pump Pneumatic Module76xx Interface Board GearmotorSystem Transducer Electronics RegulatorDome Regulator Test PortRuska Installation Unpacking the RuskaPneumatic Connections Powering Up the RuskaDrive Air Supply Port Numeric Keypad Local OperationLocal Operation Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Tare Mode Selecting Mode of OperationSelecting Language Controlling Pressure Setting the Pressure Set-PointMain Menu Triple Range OptionalVent Step/JogMenu Setup MenuLow Limit Menu Setup LimitsHigh Limit Auto VentControl Band Slew Rate AccessMenu Setup User Slew LimitReady Tolerance Step SizeBar Graph Maximum Gas Head Pressure CorrectionChanging the Number of Decimals Menu Setup UnitsPressure Filter Key Click13. Menu Setup Units Edit Name Menu Gpib Address Menu Setup RemoteMenu Setup System ProtocolDate/Time ResetCalibration Password Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMaxTime Mode of EntryEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Capabilities Remote OperationRemote Operation IEEE-488Remote/Local Operation RS-232Device Messages ConfigurationScpi Command Format Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Ruska 7350’s Full Scale Rating MaintenancePreventive Maintenance Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingRuska 7350 Pump Lubrication CleaningFuse Disconnection Instructions Preparation for Storage and ShipmentPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range