Fluke 7350 specifications Interface Panel Emulation

Page 60

RUSKA 7350

Users Manual

Questionable Status (QUES:EVENT, QUES:CONDITION, QUES:ENABLE)

Bit 0 Voltage is questionable. Set when supply voltages are not within 5%. Bit 1 Current is questionable. Always 0.

Bit 2 Time is questionable. Set when the clock has not been set.

Bit 3. Temperature is questionable. Set when the oven temperature is not within range. Bit 5. Phase is questionable. Always 0.

Bit 6 Modulation is questionable. Always 0.

Bit 7 Calibration is questionable. Set when the unit has not been calibrated. Bit 8 Pressure is questionable. Set when the pressure is overranged.

Bit 9 Always 0.

Bit 10 Always 0.

Bit 11 Always 0.

Bit 12 Always 0.

Bit 13 Instument Summary bit. Always 0.

Bit 14 Command Warning. Set whenever a command ignores a parameter. Bit 15. 0.

6005 Interface Panel Emulation

The RUSKA 7350 may be configured to emulate the IEEE-488 command set of the RUSKA Single Channel Interface Panel (Models 6005-701 and 6005-761). See the Interface Panel manual for a description of the protocol. The RUSKA 7350 emulation has the following differences:

1.The RUSKA 7350 is always in remote mode (Byte 0, Bit 2)

2.All TI strip outputs read as OFF (zero) and must be written as OFF.

3.No special functions are implemented.

4.Any message written to the RUSKA 7350 which starts with a colon as the first character is interpreted as a SCPI command.

To change from SCPI to Interface Panel Emulation via the remote interface, send the following message:

:SYSTem:LANGuage “6000”

To change from Interface Panel Emulation to SCPI via the remote interface, send the following message:

:SYSTem:LANGuage “SCPI”

5-8

Image 60
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii How to Contact Fluke General InformationIntroduction General InformationSafety Information Symbols Used in this Manual FeaturesPower-On Self Test Automatic Zero AdjustEase of Operation Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Power Supply Theory of OperationElectronics Section Back Plane BoardRuska 7350 Block Diagram IEEE-488 Interface Front PanelMotor Controller Pneumatic Module 76xx Interface BoardPneumatic Pump GearmotorElectronics Regulator Dome RegulatorSystem Transducer Test PortRuska Installation Unpacking the RuskaPowering Up the Ruska Pneumatic ConnectionsDrive Air Supply Port Local Operation Local OperationNumeric Keypad Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Mode of Operation Tare ModeSelecting Language Setting the Pressure Set-Point Main MenuControlling Pressure Triple Range OptionalVent Step/JogMenu Setup MenuMenu Setup Limits High LimitLow Limit Auto VentControl Band Access Menu Setup UserSlew Rate Slew LimitStep Size Bar Graph MaximumReady Tolerance Gas Head Pressure CorrectionMenu Setup Units Pressure FilterChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup Remote Menu Setup SystemGpib Address ProtocolReset Calibration PasswordDate/Time Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramMode of Entry MaxTimeEntering a New Program 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Remote Operation Remote OperationCapabilities IEEE-488Remote/Local Operation RS-232Configuration Scpi Command FormatDevice Messages Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Maintenance Preventive MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingCleaning Ruska 7350 Pump LubricationFuse Preparation for Storage and Shipment Disconnection InstructionsPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range