Local Operation 4

Menu

gld26.bmp

Figure 4-24. Menu Test Sweep - Menu

4.Press Run [F2] to initiate the test.

5.To pause the program, press Pause [F3]. Pause will now be highlighted and the Calibrator will continue controlling to the current set-point. The Calibrator will maintain control of the current set-point until further instructions from the user. Press Continue [F4] to resume the program.

6.To stop the program, press Stop [F5]. The program will stop running but the Calibrator will continue controlling to the current set-point.

Menu Test — Self Test

The Calibrator can perform electrical and pneumatic self test to assist in trouble shooting potential problems. Please refer to Chapter 6, Maintenance, for more details.

Note

In order to perform a pneumatic self test, the supply port must be connected to a properly set supply pressure and the test port not open to atmosphere. For absolute units, a vacuum supply must also be connected to the exhaust port.

Menu Test — Remote Test

The Menu Test Remote section of the system is used to perform a number of the diagnostic functions on the remote interface. It can be used to display the transmitted and received messages that are sent across the interface. This can be a powerful tool to assist in identifying the source of communication problems.

The Menu Test Remote menu displays information from the IEEE-488 interface. Pressing the Serial 1 [F2] or Serial 2 [F3] keys will display information from the serial ports.

4-25

Image 49
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeIntroduction General InformationSafety Information Features Symbols Used in this ManualAutomatic Zero Adjust Power-On Self TestEase of Operation Easily ProgrammableRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Theory of Operation Power SupplyElectronics Section Back Plane BoardRuska 7350 Block Diagram Front Panel IEEE-488 InterfaceMotor Controller 76xx Interface Board Pneumatic ModulePneumatic Pump GearmotorDome Regulator Electronics RegulatorSystem Transducer Test PortRuska Unpacking the Ruska InstallationPneumatic Connections Powering Up the RuskaDrive Air Supply Port Local Operation Local OperationNumeric Keypad Function KeysMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Tare Mode Selecting Mode of OperationSelecting Language Main Menu Setting the Pressure Set-PointControlling Pressure Triple Range OptionalStep/Jog VentMenu Menu SetupHigh Limit Menu Setup LimitsLow Limit Auto VentControl Band Menu Setup User AccessSlew Rate Slew LimitBar Graph Maximum Step SizeReady Tolerance Gas Head Pressure CorrectionPressure Filter Menu Setup UnitsChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup System Menu Setup RemoteGpib Address ProtocolCalibration Password ResetDate/Time Menu CalibratePreparing to Program Menu Program Storing a Sequence in MemoryMaxTime Mode of EntryEntering a New Program 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska Remote Operation Remote OperationCapabilities IEEE-488RS-232 Remote/Local OperationScpi Command Format ConfigurationDevice Messages Scpi Response FormatScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Preventive Maintenance MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsRuska 7350 Pump Lubrication CleaningFuse Disconnection Instructions Preparation for Storage and ShipmentPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of SpecificationsSpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range