Fluke specifications Ruska 7350 Pump Lubrication, Fuse, Cleaning

Page 68

RUSKA 7350

Users Manual

RUSKA 7350 Pump Lubrication

W Caution

The threaded portion of the RUSKA 7350 pump spindle requires periodic lubrication. Failure to lubricate the pump spindle can cause premature failure of the pump spindle and nut resulting in permanent pump damage.

Approved Lubrications

Jet Lube AP-5 Grease (part number 45-342)

Jet Lube MP-50 Moly-paste

Frequency of Lubrication

The frequency of lubrication is dependant upon the frequency of instrument use. On a highly utilized system, the pump must be lubricated every other week. Also the pump should be lubricated any time that the pump shows signs of loss of lubrication along the face of the pump threads.

If a pump position error occurs, along with checking the condition of the pump and pump over-travel limit switches, please check the pump lubrication. An inadequately lubricated pump can sometimes be the cause of pump-position error detection by the system software.

Lubrication Procedure

The threaded portion of the pump plunger requires periodic lubrication. The pump may be lubricated by either injecting the lubricant through the grease fitting located on the left side of the instrument or by applying the grease directly to the threads of the pump plunger. Either way, the power to the system should be removed when applying the lubrication to the pump.

When applying grease directly to the pump, extend the pump plunger such that the majority of the threads are exposed. Using the lubrication provided with the system, or one of the lubrications noted in the approved lubrication list, disperse a generous amount of lubrication all along the threaded portion of the pump plunger.

W Caution

Only the lubrication defined on the approved lubrication list above can be used. Substituting this lubrication can cause permanent damage to the pumping mechanism.

Fuse

The main fuse is a 6 AMP, 250V type 3AB-Slow blow (part number 3926870).

Cleaning

When necessary, clean externally using a damp lint-free cloth and mild liquid detergent.

6-6

Image 68
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii How to Contact Fluke General InformationIntroduction General InformationSafety Information Symbols Used in this Manual FeaturesPower-On Self Test Automatic Zero AdjustEase of Operation Easily ProgrammableStandard Equipment & Options Ruska 7350 Options List Part NumberRuska Power Supply Theory of OperationElectronics Section Back Plane BoardRuska 7350 Block Diagram Motor Controller IEEE-488 InterfaceFront Panel Pneumatic Module 76xx Interface BoardPneumatic Pump GearmotorElectronics Regulator Dome RegulatorSystem Transducer Test PortRuska Installation Unpacking the RuskaDrive Air Supply Port Powering Up the RuskaPneumatic Connections Local Operation Local OperationNumeric Keypad Function KeysCancel, Previous Menu TreeTutorial Main MenuMain Menu Entering Pressure Selecting Language Selecting Mode of OperationTare Mode Setting the Pressure Set-Point Main MenuControlling Pressure Triple Range OptionalVent Step/JogMenu Setup MenuMenu Setup Limits High LimitLow Limit Auto VentControl Band Access Menu Setup UserSlew Rate Slew LimitStep Size Bar Graph MaximumReady Tolerance Gas Head Pressure CorrectionMenu Setup Units Pressure FilterChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup Remote Menu Setup SystemGpib Address ProtocolReset Calibration PasswordDate/Time Menu CalibrateMenu Program Storing a Sequence in Memory Preparing to ProgramEntering a New Program Mode of EntryMaxTime 17. Menu Program Edit Menu Automatically Generating a Program 18. Menu Program Edit Auto MenuChanging the Name of a Program Changing an Existing ProgramChanging the Configuration Stored with a Program 20. Menu Program Edit Menu21. Menu Program Config Menu Running a ProgramMenu Test Menu Test Sweep TestMenu Test Self Test Menu Test Remote TestMenu Test Shop1 26. Menu Test Shop1 MenuMenu Display Menu Display BlankRuska Remote Operation Remote OperationCapabilities IEEE-488Remote/Local Operation RS-232Configuration Scpi Command FormatDevice Messages Scpi Response FormatANSI/IEEE 488.2-1987 Command Summary Scpi Command SummaryMODE? Example Scpi Commands ListScpi Status Registers Interface Panel Emulation Serial Operation Ruska Maintenance Preventive MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberProcessor Battery CalibrationCalibration Instructions Single Sensor PreparationStoring the Coefficients Editing the Calibration Coefficients ZeroingFuse CleaningRuska 7350 Pump Lubrication Packing Instructions Preparation for Storage and ShipmentDisconnection Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Shipping Instructions Packing the Calibrator7350 Summary of Specifications AccuracySpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range