Remote Operation 5

Configuration

Configuration

The remote interface is configured using the local interface before the remote is connected. The parameters needed varies with the interface used.

IEEE-488

Address, Protocol

RS-232

Baud Rate, Data Bits, Parity, Stop Bits

RS-485

Address, Baud Rate, Data Bits, Parity, Stop Bits

To configure the remote interface follow these steps:

1.The remote interface is configured from the Setup Remote Menu. From the Main Menu (press Prev. until the Main Menu appears) press Menu [F6] ⎢Setup [F2] ⎢Remote [F4].

2.Use the knob to highlight the desired parameter.

3.Use the numeric keypad to enter the address. The ENTER key must be pressed to change any parameter.

4.Repeat steps 2 and 3 to set all parameters needed.

Device Messages

SCPI Command Format

SCPI mnemonics have two forms: long and short. The short form is all in capital letters. The long form is the entire mnemonic. Commands may use either the short form or the entire long form. No other forms are accepted. SCPI ignores case: uppercase and lowercase are equivalent.

A SCPI command is made by following the command tree as presented in the command summary. Each level adds a mnemonic to the command separated by colons. Mnemonics enclosed in square brackets are optional and may be omitted.

Some mnemonics are followed by an optional numeric suffix. If omitted, the suffix defaults to 1.

Multiple commands may be placed in a single message separated by semicolons. Each command starts at the same level of tree where the last command stopped unless the command starts with a colon. The first command in a message and any commands starting with a colon start of the root of the command tree. IEEE 488.2 commands may occur between SCPI commands without affecting the tree level.

Command parameters are separated from the command name by one or more spaces. Multiple parameters are separated by commas. SCPI accepts numeric parameters with optional sign, decimal point, and exponent. OFF is equivalent to zero and ON is equivalent to one. Floating point numbers are rounded to the nearest integer for commands accepting integer values only.

A message is terminated by a Line Feed (hexadecimal 0A). Carriage Returns, Tabs, and other control characters are ignored.

SCPI Response Format

Only commands ending in a question mark have responses. Multiple values from a single command are separated by commas. Responses from different commands in the same message are separated by semi-colons. The response message is terminated by a Line Feed (hexadecimal 0A).

Integer responses are returned as one or more digits. Boolean values (ON and OFF values) are always returned as numbers with zero for OFF and one for ON. Floating point values are returned in the format “+d.ddddddddE+dd”.

5-3

Image 55
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeGeneral Information IntroductionSafety Information Features Symbols Used in this ManualEasily Programmable Power-On Self TestAutomatic Zero Adjust Ease of OperationRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Back Plane Board Power SupplyTheory of Operation Electronics SectionRuska 7350 Block Diagram Front Panel IEEE-488 InterfaceMotor Controller Gearmotor Pneumatic Module76xx Interface Board Pneumatic PumpTest Port Electronics RegulatorDome Regulator System TransducerRuska Unpacking the Ruska InstallationPneumatic Connections Powering Up the RuskaDrive Air Supply Port Function Keys Local OperationLocal Operation Numeric KeypadMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Tare Mode Selecting Mode of OperationSelecting Language Triple Range Optional Setting the Pressure Set-PointMain Menu Controlling PressureStep/Jog VentMenu Menu SetupAuto Vent Menu Setup LimitsHigh Limit Low LimitControl Band Slew Limit AccessMenu Setup User Slew RateGas Head Pressure Correction Step SizeBar Graph Maximum Ready ToleranceKey Click Menu Setup UnitsPressure Filter Changing the Number of Decimals13. Menu Setup Units Edit Name Menu Protocol Menu Setup RemoteMenu Setup System Gpib AddressMenu Calibrate ResetCalibration Password Date/TimePreparing to Program Menu Program Storing a Sequence in MemoryMaxTime Mode of EntryEntering a New Program 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska IEEE-488 Remote OperationRemote Operation CapabilitiesRS-232 Remote/Local OperationScpi Response Format ConfigurationScpi Command Format Device MessagesScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Observing the Software Version Number MaintenancePreventive Maintenance Observing the Ruska 7350’s Full Scale RatingCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsRuska 7350 Pump Lubrication CleaningFuse Disconnection Instructions Preparation for Storage and ShipmentPacking Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of SpecificationsSpecifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range