Fluke 7350 specifications Summary of Error Messages

Page 77

Appendix B

Summary of Error Messages

Summary of Error Messages

Negative error numbers are from the Standard Commands for Programmable Instruments (Version 1991.0).

 

Table B-1. Summary of Error Messages

 

 

Value

Description and Corrective Action

 

 

0

No Error.

 

 

-103

Invalid Separator. Check punctuation in the SCPI command.

 

 

-104

Data Type. The type of parameter data is incorrect. Check for numeric versus string

 

data.

 

 

-109

Missing Parameter. No valid parameter was found for the SCPI command.

 

 

-110

Command Header. The command name is not valid.

 

 

-113

Command Unknown. The command specified does not exist.

 

 

-114

Header Suffix. The numeric suffix for the command name is out of range.

 

 

-201

Remote Required. Remote command conflicted with local interface. Return local

 

interface to main Menu.

 

 

-221

Settings Conflict. The command could not be executed due to the current state of

 

the Calibrator. Some commands cannot be executed while a program, self-test, or

 

calibration is in progress.

 

 

-222

Out of Range. The value is not within the valid range. For pressures, check your

 

high and low limits.

 

 

-281

Cannot create program. Program memory is full.

 

 

-282

Illegal Program Name. The name specified is not valid or does not exist.

 

 

-284

Program Currently Running. The command cannot be executed while a program is

 

running.

 

 

-285

Program Syntax Error. The syntax of the program definition is not correct.

 

 

B-1

Image 77
Contents Ruska Limited Warranty and Limitation of Liability Table of Contents Theory of Operation Remote Operation Preparation for Storage and Shipment List of Tables Ruska List of Figures Viii General Information How to Contact FlukeIntroduction General InformationSafety Information Features Symbols Used in this ManualAutomatic Zero Adjust Power-On Self TestEase of Operation Easily ProgrammableRuska 7350 Options List Part Number Standard Equipment & OptionsRuska Theory of Operation Power SupplyElectronics Section Back Plane BoardRuska 7350 Block Diagram Motor Controller IEEE-488 InterfaceFront Panel 76xx Interface Board Pneumatic ModulePneumatic Pump GearmotorDome Regulator Electronics RegulatorSystem Transducer Test PortRuska Unpacking the Ruska InstallationDrive Air Supply Port Powering Up the RuskaPneumatic Connections Local Operation Local OperationNumeric Keypad Function KeysMenu Tree Cancel, PreviousMain Menu TutorialMain Menu Entering Pressure Selecting Language Selecting Mode of OperationTare Mode Main Menu Setting the Pressure Set-PointControlling Pressure Triple Range OptionalStep/Jog VentMenu Menu SetupHigh Limit Menu Setup LimitsLow Limit Auto VentControl Band Menu Setup User AccessSlew Rate Slew LimitBar Graph Maximum Step SizeReady Tolerance Gas Head Pressure CorrectionPressure Filter Menu Setup UnitsChanging the Number of Decimals Key Click13. Menu Setup Units Edit Name Menu Menu Setup System Menu Setup RemoteGpib Address ProtocolCalibration Password ResetDate/Time Menu CalibratePreparing to Program Menu Program Storing a Sequence in MemoryEntering a New Program Mode of EntryMaxTime 17. Menu Program Edit Menu 18. Menu Program Edit Auto Menu Automatically Generating a ProgramChanging an Existing Program Changing the Name of a Program20. Menu Program Edit Menu Changing the Configuration Stored with a ProgramRunning a Program 21. Menu Program Config MenuMenu Test Sweep Test Menu TestMenu Test Remote Test Menu Test Self Test26. Menu Test Shop1 Menu Menu Test Shop1Menu Display Blank Menu DisplayRuska Remote Operation Remote OperationCapabilities IEEE-488RS-232 Remote/Local OperationScpi Command Format ConfigurationDevice Messages Scpi Response FormatScpi Command Summary ANSI/IEEE 488.2-1987 Command SummaryMODE? List Example Scpi CommandsScpi Status Registers Interface Panel Emulation Serial Operation Ruska Preventive Maintenance MaintenanceObserving the Ruska 7350’s Full Scale Rating Observing the Software Version NumberCalibration Processor BatteryPreparation Calibration Instructions Single SensorStoring the Coefficients Zeroing Editing the Calibration CoefficientsFuse CleaningRuska 7350 Pump Lubrication Packing Instructions Preparation for Storage and ShipmentDisconnection Instructions Fluke Calibration Westpark Drive HOUSTON, TX USA Packing the Calibrator Shipping Instructions7350 Accuracy Summary of Specifications Specifications Summary of Specifications Ruska Summary of Error Messages Summary of Error MessagesSetpoint is out of range