Appendix B
Summary of Error Messages
Summary of Error Messages
Negative error numbers are from the Standard Commands for Programmable Instruments (Version 1991.0).
| Table |
|
|
Value | Description and Corrective Action |
|
|
0 | No Error. |
|
|
Invalid Separator. Check punctuation in the SCPI command. | |
|
|
Data Type. The type of parameter data is incorrect. Check for numeric versus string | |
| data. |
|
|
Missing Parameter. No valid parameter was found for the SCPI command. | |
|
|
Command Header. The command name is not valid. | |
|
|
Command Unknown. The command specified does not exist. | |
|
|
Header Suffix. The numeric suffix for the command name is out of range. | |
|
|
Remote Required. Remote command conflicted with local interface. Return local | |
| interface to main Menu. |
|
|
Settings Conflict. The command could not be executed due to the current state of | |
| the Calibrator. Some commands cannot be executed while a program, |
| calibration is in progress. |
|
|
Out of Range. The value is not within the valid range. For pressures, check your | |
| high and low limits. |
|
|
Cannot create program. Program memory is full. | |
|
|
Illegal Program Name. The name specified is not valid or does not exist. | |
|
|
Program Currently Running. The command cannot be executed while a program is | |
| running. |
|
|
Program Syntax Error. The syntax of the program definition is not correct. | |
|
|